Comparison of optic disc topography measured by Retinal Thickness Analyzer with measurement by Heidelberg Retina Tomograph II
- PMID: 12738557
- DOI: 10.1016/s0021-5155(02)00699-8
Comparison of optic disc topography measured by Retinal Thickness Analyzer with measurement by Heidelberg Retina Tomograph II
Abstract
Purpose: To compare the optic disc topography measurements from a digitized laser slit-lamp and a confocal scanning laser ophthalmoscope (SLO).
Methods: Ten normal subjects (10 eyes) were recruited. Topographic measurements of the optic disc were performed three times in all eyes using the Retinal Thickness Analyzer (RTA) and the Heidelberg Retina Tomograph II (HRT II), and the mean values determined. The mean values of 11 optic disc parameters were compared between the two instruments by the Wilcoxon signed rank test. To test the reproducibility of the topographic measurements, the coefficients of variation (CV) of the topographic parameters among the three measurements from each of the two instruments were compared using the Wilcoxon signed rank test.
Results: The mean cup depth, mean retinal nerve fiber layer (RNFL) thickness, and RNFL cross-sectional area were significantly smaller when measured by RTA compared with HRT II (P =.0067, P =.0364, P =.0467, respectively). The mean CVs from the RTA were larger than those from the HRT II for all parameters; however, the differences in all parameters did not reach significance.
Conclusion: The RTA measured smaller z-axis values compared with the HRT II. The reproducibility of the topographic data was not significantly different between the instruments.
Similar articles
-
Comparison of optic nerve head topography in healthy adults using a Heidelberg retina tomograph and retinal thickness analyzer.Int Ophthalmol. 2007 Feb;27(1):1-9. doi: 10.1007/s10792-007-9051-z. Epub 2007 Apr 11. Int Ophthalmol. 2007. PMID: 17426923
-
Assessment of optic nerve head topographic parameters with a confocal scanning laser ophthalmoscope.Clin Exp Ophthalmol. 2004 Jun;32(3):259-64. doi: 10.1111/j.1442-9071.2004.00790.x. Clin Exp Ophthalmol. 2004. PMID: 15180837
-
[Correlations between retinal thickness analyzer (RTA) and confocal scanning laser tomography (HRT) in optic disc analysis].Arch Soc Esp Oftalmol. 2004 Jan;79(1):21-5. doi: 10.4321/s0365-66912004000100006. Arch Soc Esp Oftalmol. 2004. PMID: 14752698 Spanish.
-
[New examination methods for macular disorders--application of diagnosis and treatment].Nippon Ganka Gakkai Zasshi. 2000 Dec;104(12):899-942. Nippon Ganka Gakkai Zasshi. 2000. PMID: 11193944 Review. Japanese.
-
Measurement of the retinal nerve fiber layer thickness in clinical routine.Curr Opin Ophthalmol. 1998 Apr;9(2):77-87. doi: 10.1097/00055735-199804000-00015. Curr Opin Ophthalmol. 1998. PMID: 10180519 Review.
Cited by
-
Alzheimer's disease pathophysiology in the Retina.Prog Retin Eye Res. 2024 Jul;101:101273. doi: 10.1016/j.preteyeres.2024.101273. Epub 2024 May 15. Prog Retin Eye Res. 2024. PMID: 38759947 Free PMC article. Review.
-
Intraobserver and interobserver agreement of computer software-assisted optic nerve head photoplanimetry.Jpn J Ophthalmol. 2014 Jan;58(1):56-61. doi: 10.1007/s10384-013-0280-6. Epub 2013 Oct 8. Jpn J Ophthalmol. 2014. PMID: 24100930
-
Comparison of optic nerve head topography in healthy adults using a Heidelberg retina tomograph and retinal thickness analyzer.Int Ophthalmol. 2007 Feb;27(1):1-9. doi: 10.1007/s10792-007-9051-z. Epub 2007 Apr 11. Int Ophthalmol. 2007. PMID: 17426923
-
An improved medical decision support system to identify the diabetic retinopathy using fundus images.J Med Syst. 2012 Dec;36(6):3573-81. doi: 10.1007/s10916-012-9833-3. Epub 2012 Mar 6. J Med Syst. 2012. PMID: 22392564
-
Repeatability and reproducibility of optic nerve head topography using the retinal thickness analyzer.Graefes Arch Clin Exp Ophthalmol. 2006 Feb;244(2):192-8. doi: 10.1007/s00417-005-0027-2. Epub 2005 Aug 5. Graefes Arch Clin Exp Ophthalmol. 2006. PMID: 16082558
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources