Computer-aided placement of endosseous oral implants in patients after ablative tumour surgery: assessment of accuracy
- PMID: 12755784
- DOI: 10.1034/j.1600-0501.2003.110812.x
Computer-aided placement of endosseous oral implants in patients after ablative tumour surgery: assessment of accuracy
Abstract
The objective of this study was to evaluate the feasibility and accuracy of a novel surgical computer-aided navigation system for the placement of endosseous implants in patients after ablative tumour surgery. Pre-operative planning was performed by developing a prosthetic concept and modifying the implant position according to surgical requirements after high-resolution computed tomography (HRCT) scans with VISIT, a surgical planning and navigation software developed at the Vienna General Hospital. The pre-operative plan was transferred to the patients intraoperatively using surgical navigation software and optical tracking technology. The patients were HRCT-scanned again to compare the position of the implants with the pre-operative plan on reformatted CT-slices after matching of the pre- and post-operative data sets using the mutual information-technique. A total of 32 implants was evaluated. The mean deviation was 1.1 mm (range: 0-3.5 mm). The mean angular deviation of the implants was 6.4 degrees (range: 0.4 degrees - 17.4 degrees, variance: 13.3 degrees ). The results demonstrate, that adequate accuracy in placing endosseous oral implants can be delivered to patients with most difficult implantologic situations.
Similar articles
-
Evaluation of accuracy of computer-aided intraoperative positioning of endosseous oral implants in the edentulous mandible.Clin Oral Implants Res. 2002 Feb;13(1):59-64. doi: 10.1034/j.1600-0501.2002.130107.x. Clin Oral Implants Res. 2002. PMID: 12005146
-
Accuracy of Dynamic Navigation for Dental Implant Placement-Model-Based Evaluation.J Oral Implantol. 2016 Oct;42(5):399-405. doi: 10.1563/aaid-joi-D-16-00025. Epub 2016 Jun 6. J Oral Implantol. 2016. PMID: 27267658
-
Transfer precision of three-dimensional implant planning with CT assisted offline navigation.Int J Comput Dent. 2008;11(3-4):213-25. Int J Comput Dent. 2008. PMID: 19216313 English, German.
-
Dynamic Navigation for Dental Implant Surgery.Oral Maxillofac Surg Clin North Am. 2019 Nov;31(4):539-547. doi: 10.1016/j.coms.2019.08.001. Oral Maxillofac Surg Clin North Am. 2019. PMID: 31563194 Review.
-
Current state of the art of computer-guided implant surgery.Periodontol 2000. 2017 Feb;73(1):121-133. doi: 10.1111/prd.12175. Periodontol 2000. 2017. PMID: 28000275 Review.
Cited by
-
Accuracy of intraoral real-time navigation versus static, CAD/CAM-manufactured pilot drilling guides in dental implant surgery: an in vitro study.Int J Implant Dent. 2022 Oct 6;8(1):41. doi: 10.1186/s40729-022-00430-6. Int J Implant Dent. 2022. PMID: 36198996 Free PMC article.
-
Oral Implant Imaging: A Review.Malays J Med Sci. 2015 May-Jun;22(3):7-17. Malays J Med Sci. 2015. PMID: 26715891 Free PMC article.
-
Guided implant surgery on oral cancer patients: in vitro study.Int J Comput Assist Radiol Surg. 2011 Jan;6(1):135-42. doi: 10.1007/s11548-010-0490-z. Epub 2010 Jul 27. Int J Comput Assist Radiol Surg. 2011. PMID: 20661657
-
Automated dental implantation using image-guided robotics: registration results.Int J Comput Assist Radiol Surg. 2011 Sep;6(5):627-34. doi: 10.1007/s11548-010-0543-3. Epub 2011 Jan 8. Int J Comput Assist Radiol Surg. 2011. PMID: 21221831
-
Accuracy of computer-assisted implant placement with insertion templates.GMS Interdiscip Plast Reconstr Surg DGPW. 2016 May 13;5:Doc15. doi: 10.3205/iprs000094. eCollection 2016. GMS Interdiscip Plast Reconstr Surg DGPW. 2016. PMID: 27274440 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical