Artifacts in T(1rho)-weighted imaging: correction with a self-compensating spin-locking pulse
- PMID: 12762988
- DOI: 10.1016/s1090-7807(02)00197-0
Artifacts in T(1rho)-weighted imaging: correction with a self-compensating spin-locking pulse
Abstract
Significant artifacts arise in T(1rho)-weighted imaging when nutation angles suffer small deviations from their expected values. These artifacts vary with spin-locking time and amplitude, severely limiting attempts to perform quantitative imaging or measurement of T(1rho) relaxation times. A theoretical model explaining the origin of these artifacts is presented in the context of a T(1rho)-prepared fast spin-echo imaging sequence. Experimentally obtained artifacts are compared to those predicted by theory and related to B(1) inhomogeneity. Finally, a "self-compensating" spin-locking preparatory pulse cluster is presented, in which the second half of the spin-locking pulse is phase-shifted by 180 degrees. Use of this pulse sequence maintains relatively uniform signal intensity despite large variations in flip angle, greatly reducing artifacts in T(1rho)-weighted imaging.