Development and testing of a pelvic goniometer designed to measure pelvic tilt and hip flexion
- PMID: 12763443
- DOI: 10.1016/s0268-0033(03)00049-4
Development and testing of a pelvic goniometer designed to measure pelvic tilt and hip flexion
Abstract
Objective: To determine the reliability and validity of a pelvic goniometer designed to measure the pelvic tilt and hip flexion during seated posture.
Background: Assessment of the seated posture requires measurement of the pelvis and hip. Determining accurate pelvic tilt and hip flexion angles during sitting is often difficult using standard techniques. A pelvic goniometer has been designed to measure pelvic tilt and hip flexion angle of persons in a seated posture.
Methods: VALIDATION of the pelvic goniometer was done radiographically. Ten male volunteers sat in three postures--erect, forward or anterior tilt, and posterior tilt. Pelvic tilt and hip angle were recorded using radiographs and the pelvic goniometer. Reliability of pelvic and conventional goniometers was done using seated nondisabled subjects with physical therapists performing measurements.
Validation: the average differences and correlation between the pelvic goniometer and radiographic measures were as follows--pelvic tilt: -4.9 degrees, 0.93; hip angle 1.2 degrees, 0.81. Reliability: average range of hip angle across three measures was about 3 degrees for both goniometers.
Conclusions: The data indicate that the pelvic goniometer has utility in measuring pelvic tilt and hip angle, especially within the seated posture. Because it measures both pelvic tilt and hip angle, the pelvic goniometer has an advantage over conventional goniometers that only measure the latter.
Relevance: A valid and reliable tool that measures pelvic tilt and hip angle of persons in a seated posture is needed for clinical research and practice. Its applications include wheelchair seating evaluations and ergonomic assessments of seated workers.
Similar articles
-
Development of a noninvasive measure of pelvic and hip angles in seated posture.Arch Phys Med Rehabil. 2002 Nov;83(11):1597-602. doi: 10.1053/apmr.2002.35468. Arch Phys Med Rehabil. 2002. PMID: 12422332
-
Measurement of the normal contralateral hip flexion angle and posterior pelvic tilt angle using a pelvic guide pin in total hip arthroplasty.Clin Biomech (Bristol). 2023 Jun;106:105982. doi: 10.1016/j.clinbiomech.2023.105982. Epub 2023 May 4. Clin Biomech (Bristol). 2023. PMID: 37207497
-
Does Degenerative Lumbar Spine Disease Influence Femoroacetabular Flexion in Patients Undergoing Total Hip Arthroplasty?Clin Orthop Relat Res. 2016 Aug;474(8):1788-97. doi: 10.1007/s11999-016-4787-2. Epub 2016 Mar 28. Clin Orthop Relat Res. 2016. PMID: 27020429 Free PMC article.
-
Intratester and intertester reliability and criterion validity of the parallelogram and universal goniometers for active knee flexion in healthy subjects.Physiother Res Int. 1997;2(3):150-66. doi: 10.1002/pri.97. Physiother Res Int. 1997. PMID: 9421820 Review.
-
Photogrammetric Assessment of Upper Body Posture Using Postural Angles: A Literature Review.J Chiropr Med. 2017 Jun;16(2):131-138. doi: 10.1016/j.jcm.2017.01.005. Epub 2017 Mar 18. J Chiropr Med. 2017. PMID: 28559753 Free PMC article. Review.
Cited by
-
The modified Thomas test is not a valid measure of hip extension unless pelvic tilt is controlled.PeerJ. 2016 Aug 11;4:e2325. doi: 10.7717/peerj.2325. eCollection 2016. PeerJ. 2016. PMID: 27602291 Free PMC article.
-
Posterior pelvic tilt is a risk factor for rectal prolapse: a propensity score matching analysis.Tech Coloproctol. 2020 May;24(5):463-469. doi: 10.1007/s10151-020-02179-9. Epub 2020 Mar 13. Tech Coloproctol. 2020. PMID: 32170509
-
Effects of anterior weight-shifting methods on sitting balance in wheelchair-dependent patients with spinal cord injury.J Phys Ther Sci. 2018 Mar;30(3):393-397. doi: 10.1589/jpts.30.393. Epub 2018 Mar 2. J Phys Ther Sci. 2018. PMID: 29581657 Free PMC article.
-
Development of valid and reliable measures of postural stability.J Spinal Cord Med. 2007;30(1):40-9. doi: 10.1080/10790268.2007.11753913. J Spinal Cord Med. 2007. PMID: 17385269 Free PMC article.
-
A Sensor-Based Screening Tool for Identifying High Pelvic Mobility in Patients Due to Undergo Total Hip Arthroplasty.Sensors (Basel). 2020 Oct 30;20(21):6182. doi: 10.3390/s20216182. Sensors (Basel). 2020. PMID: 33143034 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials