CSF flow imaging in the management of third ventriculostomy with a reversed fast imaging with steady-state precession sequence
- PMID: 12764663
- DOI: 10.1007/s00330-002-1776-5
CSF flow imaging in the management of third ventriculostomy with a reversed fast imaging with steady-state precession sequence
Abstract
The aim of this study was evaluation of a fast and slow-flow sensitive 2D steady-state free-precession sequence for its capability to prove the patency of endoscopic third ventriculostomy (TVS) in obstructive hydrocephalus, and to exclude communicating third ventricle prior to TVS. We compared gated and ungated variants of this sequence for this purpose. Twenty-three patients with obstructive hydrocephalus underwent 36 MR examinations with a 2D reversed fast imaging with steady-state precession (PSIF) sequence in a retrospectively cardiac gated (cine) and a faster but ungated version beside T1- and T2-weighted sequences in three planes. Thirteen patients were examined both before and after TVS, 4 patients solely before, and 6 patients solely after TVS. Imaging diagnoses were compared with intraoperative findings and clinical findings after TVS. Preoperative diagnosis of non-communicating third ventricle and cisterns was intraoperatively confirmed in 16 of 17 cases. Preoperative MRI was inconclusive in 1 case. Postoperative MRI revealed sufficient TVS in 16 of 19 cases and obstructed TVS in 3 of 19 cases due to several reasons. Findings at MRI were consistent in 19 of 19 cases with the clinical course and intraoperative results. The faster but ungated PSIF sequence was found to be diagnostically equivalent to the cardiac gated cine sequence. The CSF flow imaging with a 2D reversed fast imaging with steady-state precession sequence in conjunction with conventional T1- and T2-weighted images is a fast and reliable tool for pre- and postoperative functional evaluation in third ventriculostomy.
Similar articles
-
Dynamic magnetic resonance imaging of endoscopic third ventriculostomy patency with differently acquired fast imaging with steady-state precession sequences.Bosn J Basic Med Sci. 2014 Aug 16;14(3):165-70. doi: 10.17305/bjbms.2014.3.37. Bosn J Basic Med Sci. 2014. PMID: 25172977 Free PMC article.
-
Three-dimensional constructive interference in steady-state magnetic resonance imaging in obstructive hydrocephalus: relevance for endoscopic third ventriculostomy and clinical results.J Neurosurg. 2008 Nov;109(5):931-8. doi: 10.3171/JNS/2008/109/11/0931. J Neurosurg. 2008. PMID: 18976087
-
Assessment of third ventriculostomy patency with the 3D-SPACE technique: a preliminary multicenter research study.J Neurosurg. 2015 Jun;122(6):1347-55. doi: 10.3171/2014.10.JNS14298. Epub 2015 Apr 10. J Neurosurg. 2015. PMID: 25859808
-
Endoscopic third ventriculostomy: anesthetic implications.Minim Invasive Neurosurg. 2004 Jun;47(3):151-3. doi: 10.1055/s-2004-818516. Minim Invasive Neurosurg. 2004. PMID: 15343430 Review.
-
Endoscopic third ventriculostomy.Neurosurg Focus. 2005 Dec 15;19(6):E11. Neurosurg Focus. 2005. PMID: 16398476 Review.
Cited by
-
Endoscopic third ventriculostomy for obstructive hydrocephalus.Neurosurg Rev. 2005 Jan;28(1):1-34; discussion 35-8. doi: 10.1007/s10143-004-0365-2. Epub 2004 Nov 27. Neurosurg Rev. 2005. PMID: 15570445 Review.
-
Natural History of Endoscopic Third Ventriculostomy in Adults: Serial Evaluation with High-Resolution CISS.AJNR Am J Neuroradiol. 2018 Dec;39(12):2231-2236. doi: 10.3174/ajnr.A5861. Epub 2018 Nov 15. AJNR Am J Neuroradiol. 2018. PMID: 30442699 Free PMC article.
-
Quantitative assessment of physiological cerebrospinal fluid flow in the cervical spinal canal with 3.0T phase-contrast cine MRI.Neural Regen Res. 2012 Jun 25;7(18):1392-7. doi: 10.3969/j.issn.1673-5374.2012.18.005. Neural Regen Res. 2012. PMID: 25657672 Free PMC article.
-
The occurrence of obstructive vs absorptive hydrocephalus in newborns and infants: relevance to treatment choices.Childs Nerv Syst. 2006 Dec;22(12):1543-63. doi: 10.1007/s00381-006-0193-5. Epub 2006 Nov 7. Childs Nerv Syst. 2006. PMID: 17091274 Review.
-
Intracranial cerebrospinal fluid spaces imaging using a pulse-triggered three-dimensional turbo spin echo MR sequence with variable flip-angle distribution.Eur Radiol. 2011 Feb;21(2):402-10. doi: 10.1007/s00330-010-1925-1. Epub 2010 Aug 21. Eur Radiol. 2011. PMID: 20725835
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical