Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography
- PMID: 12766052
- DOI: 10.1167/iovs.02-0731
Precorneal and pre- and postlens tear film thickness measured indirectly with optical coherence tomography
Abstract
Purpose: To demonstrate the feasibility of indirectly measuring the precorneal tear film thickness and pre- and postlens tear film (PLTF) thickness using optical coherence tomography (OCT).
Methods: Central corneal thickness (C(1)) which includes the tear film (T) of both eyes of 40 non-contact lens wearers was measured using OCT after calibration. The mean age of the 40 subjects was 31.2 +/- 9.3 years with a mean horizontal K-reading of 7.87 mm. Rigid contact lenses with base curves 0.3- to 0.5-mm steeper than the flattest K of the eye were fitted to measure real corneal thickness (C(2)), independently of the postlens tear film. T was calculated by T = C(1) - C(2). To measure pre- and postlens tear film thickness, Focus Night & Day and Acuvue lenses (Vistakon, Johnson & Johnson Vision Care, Jacksonville, FL) were fitted on both eyes. Central soft lens thickness (L(1)), which includes the prelens tear film (P), was measured by OCT in situ and in saline in a wet cell (L(2)). P was calculated by P = L(1) - L(2). Thickness of the central cornea plus the postlens tear film (C(3)) was measured during lens wearing. Postlens tear film (PLTF) was calculated by PLTF = C(3) - C(2). RESULTS. The mean +/- SD precorneal tear film thickness was 3.3 +/- 1.5 microm (range, 0-6.9) before lens insertion and 4.7 +/- 2.3 microm (range, 0.7-11.0) after lens fitting, which was significantly thicker (paired t-test: P < 0.01). The prelens tear film thickness was 3.9 +/- 2.6 and 3.6 +/- 2.1 microm (mean +/- SD; paired t-test: P = 0.52) and the postlens tear film thickness was 4.5 +/- 2.3 and 4.7 +/- 3.1 microm (paired t-test: P = 0.08) on and under Focus Night & Day and Acuvue lenses, respectively. Post hoc tests showed that precorneal (baseline) and prelens tear films were equivalent, and each was different (thinner; Tukey honestly significant difference P < 0.05) from the postlens tear film.
Conclusions: OCT can noninvasively measure the thickness of the precorneal and prelens tear film as well as the postlens tear film. The thickness of the normal precorneal tear film is approximately 3 micro m and becomes thicker after lens fitting. The postlens tear film is thicker than the precorneal and prelens tear films with soft contact lenses. The thickness of both pre- and postlens tear films appears to be independent of the investigated lens types.
Similar articles
-
Thickness of the pre- and post-contact lens tear film measured in vivo by interferometry.Invest Ophthalmol Vis Sci. 2003 Jan;44(1):68-77. doi: 10.1167/iovs.02-0377. Invest Ophthalmol Vis Sci. 2003. PMID: 12506057
-
Thinning rate of the precorneal and prelens tear films.Invest Ophthalmol Vis Sci. 2005 Jul;46(7):2353-61. doi: 10.1167/iovs.05-0094. Invest Ophthalmol Vis Sci. 2005. PMID: 15980222
-
The impact of contact lens care solutions on the thickness of the tear film and contact lens.Cornea. 2005 Oct;24(7):825-32. doi: 10.1097/01.ico.0000157904.52305.29. Cornea. 2005. PMID: 16160499 Clinical Trial.
-
OCT applications in contact lens fitting.Cont Lens Anterior Eye. 2022 Aug;45(4):101540. doi: 10.1016/j.clae.2021.101540. Epub 2021 Nov 17. Cont Lens Anterior Eye. 2022. PMID: 34799247 Review.
-
Tear film-contact lens interactions.Adv Exp Med Biol. 1994;350:403-10. doi: 10.1007/978-1-4615-2417-5_70. Adv Exp Med Biol. 1994. PMID: 8030509 Review.
Cited by
-
Nanocarriers for ocular drug delivery: Recent advances and future opportunities.Indian J Ophthalmol. 2023 Jun;71(6):2355-2366. doi: 10.4103/ijo.IJO_1893_22. Indian J Ophthalmol. 2023. PMID: 37322644 Free PMC article. Review.
-
Tear Film Changes and Ocular Symptoms Associated with Soft Contact Lens Wear.Vision (Basel). 2025 Apr 1;9(2):27. doi: 10.3390/vision9020027. Vision (Basel). 2025. PMID: 40265395 Free PMC article.
-
Ophthalmic Hydrogel Contact Lens Material Containing Magnesium Oxide Nanoparticles and 3-(Trifluoromethyl)styrene for Biomedical Application.Micromachines (Basel). 2022 Nov 2;13(11):1897. doi: 10.3390/mi13111897. Micromachines (Basel). 2022. PMID: 36363917 Free PMC article.
-
An imaging system integrating optical coherence tomography and interferometry for in vivo measurement of the thickness and dynamics of the tear film.Biomed Eng Online. 2018 Nov 1;17(1):164. doi: 10.1186/s12938-018-0597-y. Biomed Eng Online. 2018. PMID: 30382929 Free PMC article.
-
Fourier-Domain OCT Imaging of the Ocular Surface and Tear Film Dynamics: A Review of the State of the Art and an Integrative Model of the Tear Behavior During the Inter-Blink Period and Visual Fixation.J Clin Med. 2020 Mar 2;9(3):668. doi: 10.3390/jcm9030668. J Clin Med. 2020. PMID: 32131486 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous