Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2003;32(5):341-5.
doi: 10.1002/lsm.10177.

Preliminary study of laser welding for aortic dissection in a porcine model using a diode laser with indocyanine green

Affiliations

Preliminary study of laser welding for aortic dissection in a porcine model using a diode laser with indocyanine green

Masanori Fujita et al. Lasers Surg Med. 2003.

Abstract

Background and objectives: The objective of this study was to determine whether a dissected aorta could be welded by a diode laser with a solder using an in vitro porcine aortic dissection model.

Study design/materials and methods: Porcine aortic strips were dissected into two flaps and the dissected faces were immersed in a solution of indocyanine green. The two flaps were pressed at 0.2 kg/cm2 with contact between the two immersed faces. The pressed flaps were irradiated with a diode laser (810 nm) at intensities of 170-425 W/cm2 for 8 seconds. The welded flaps were studied by light microscopy and the adhesive strengths were measured.

Results: The irradiated flaps were successfully welded. The breaking stress, the maximum stress recorded in a stress-strain curve, increased with increase in irradiation intensity up to 396 W/cm2 (2.7 x 10(2) mmHg) and decreased when the intensity reached 425 W/cm2. In the specimen irradiated at 396 W/cm2, the welded faces showed continuous fusion of elastin layers, while some voids were seen between the welded faces in the specimen irradiated at 425 W/cm2.

Conclusions: The dissected porcine aortas were successfully welded using a laser with solder. The results suggest that the welded aorta can bear physiological blood pressure.

PubMed Disclaimer