The effect of implant overlap on the mechanical properties of the femur
- PMID: 12777906
- DOI: 10.1097/01.TA.0000060999.54287.39
The effect of implant overlap on the mechanical properties of the femur
Abstract
Background: The most biomechanically stable relationship between the side plate of a compression hip screw (CHS) and retrograde intramedullary (IM) femoral nail has not been described in the literature. This becomes a clinical issue when treating supracondylar femur fractures with a retrograde nail in patients with a history of compression hip screw fixation of intertrochanteric fractures. The proximal end of the nail and the interlocking screws may act as a stress riser in the femoral diaphysis. The purpose of this study is to determine the biomechanical consequences of the amount of implant overlap between a CHS plate and retrograde IM femoral nail.
Methods: Nine paired fresh-frozen cadaver femora from elderly donors were cleaned of soft tissue and fixed with uniaxial strain gauges. Each femur was loaded three times in a fall-loading configuration to 50 kg at a rate of 1 Hz. The study consisted of two phases. In phase 1, six pair were randomly divided into a control and test femur from each pair. Three states were compared on each test femur: uninstrumented, instrumented with CHS, and instrumented with CHS and retrograde nail. The control femur consisted of the matched femur tested in two states: uninstrumented and instrumented with a CHS. The femora were then loaded to failure. The tests were performed with the retrograde nail and CHS gapped 3 cm, kissing, and overlapping by two screw holes (two pair for each state). In phase 2, each of the remaining three pair were instrumented with a CHS and retrograde nail overlapping in one femur and gapped in the matched femur and tested in the same manner. Data analysis was performed using Pearson's correlation coefficients between groups. Paired samples t tests were used to compare differences within test states and independent samples t tests were used to compare differences between femora. Mean strain at 50-kg load, load-versus-strain patterns, failure patterns, and load and strain at failure were recorded. RESULTS; Correlation coefficients were greater than 0.98 within and between pairs (p < 0.001). There were statistically significant differences (p < 0.05) in strain patterns between the uninstrumented, CHS, and CHS/IM test states. The addition of a side plate significantly (p < 0.05) increased lateral compressive strains in the femoral diaphysis. Mean strain at 50-kg load was significantly (p < 0.05) altered by the addition of the retrograde nail in all three implant orientations. Gapped implants failed at lower loads and strains than femurs with kissing and overlapping implants. Gapped constructs failed at lower loads than control states. Overlapped constructs tolerated the highest loads and strains before failure.
Conclusion: Strain patterns are altered by the degree of implant overlap in the proximal femoral diaphysis. Femora with uninstrumented intervals between retrograde nails and side plates fail at lower loads than femora without retrograde nails and those with kissing or overlapping implants. Kissing or overlapping instrumentation increases load to failure and creates a more biomechanically stable construct than gapped implants. The findings of this study suggest an overlapping implant orientation in the femur increases failure load at the implant interface.
Similar articles
-
Distal femoral fixation: a biomechanical comparison of trigen retrograde intramedullary (i.m.) nail, dynamic condylar screw (DCS), and locking compression plate (LCP) condylar plate.J Trauma. 2009 Feb;66(2):443-9. doi: 10.1097/TA.0b013e31815edeb8. J Trauma. 2009. PMID: 19204519
-
Biomechanical comparison of two side plate fixation techniques in an unstable intertrochanteric osteotomy model: Sliding Hip Screw and Percutaneous Compression Plate.Clin Biomech (Bristol). 2007 Dec;22(10):1112-8. doi: 10.1016/j.clinbiomech.2007.07.016. Epub 2007 Sep 27. Clin Biomech (Bristol). 2007. PMID: 17900766
-
Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques.J Orthop Trauma. 2007 Sep;21(8):544-8. doi: 10.1097/BOT.0b013e31814b822e. J Orthop Trauma. 2007. PMID: 17805021
-
The geriatric distal femur fracture: nail, plate or both?Eur J Orthop Surg Traumatol. 2023 Jul;33(5):1485-1493. doi: 10.1007/s00590-022-03337-5. Epub 2022 Jul 27. Eur J Orthop Surg Traumatol. 2023. PMID: 35895117 Review.
-
Treatment of Distal Femur Fractures with a Combined Nail-Plate Construct: Techniques and Outcomes.J Long Term Eff Med Implants. 2021;31(3):15-26. doi: 10.1615/JLongTermEffMedImplants.2021038016. J Long Term Eff Med Implants. 2021. PMID: 34369718 Review.
Cited by
-
Minimally invasive dual plating of a complex femoral fracture; a case report.Trauma Case Rep. 2023 Jan 9;43:100768. doi: 10.1016/j.tcr.2023.100768. eCollection 2023 Feb. Trauma Case Rep. 2023. PMID: 36686415 Free PMC article.
-
Plate on plate osteosynthesis for the treatment of nonhealed periplate fractures.ISRN Orthop. 2014 Mar 11;2014:367490. doi: 10.1155/2014/367490. eCollection 2014. ISRN Orthop. 2014. PMID: 24967127 Free PMC article.
-
Periprosthetic fracture management around total knee arthroplasty.J Orthop. 2021 Jan 5;23:239-245. doi: 10.1016/j.jor.2020.12.024. eCollection 2021 Jan-Feb. J Orthop. 2021. PMID: 33613007 Free PMC article. No abstract available.
-
Osteoporotic Distal Femur Fractures: An Overview.Indian J Orthop. 2025 Feb 12;59(3):311-325. doi: 10.1007/s43465-025-01345-1. eCollection 2025 Mar. Indian J Orthop. 2025. PMID: 40201919 Review.
-
[Retrograde intramedullary nailing for periprosthetic fractures of the distal femur].Oper Orthop Traumatol. 2014 Oct;26(5):438-54. doi: 10.1007/s00064-014-0303-6. Epub 2014 Sep 13. Oper Orthop Traumatol. 2014. PMID: 25213461 Clinical Trial. German.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical