Comparison of microscan broth microdilution, synergy quad plate agar dilution, and disk diffusion screening methods for detection of high-level aminoglycoside resistance in enterococcus species
- PMID: 12791910
- PMCID: PMC156517
- DOI: 10.1128/JCM.41.6.2703-2705.2003
Comparison of microscan broth microdilution, synergy quad plate agar dilution, and disk diffusion screening methods for detection of high-level aminoglycoside resistance in enterococcus species
Abstract
We compared the dried MicroScan microdilution panel, Synergy Quad plate agar dilution, and high-potency disk diffusion screening methods for the detection of high-level aminoglycoside resistance in 815 enterococcal bloodstream isolates. Agreement between the three methods was 99% when testing for high-level gentamicin resistance and 96% when testing for high-level streptomycin resistance.
References
-
- Chen, Y.-S., S. A. Marshall, P. L. Winokur, S. L. Coffman, W. W. Wilkie, P. R. Murray, C. A. Spiegel, M. A. Pfaller, G. V. Doern, and R. N. Jones. 1998. Use of molecular and reference susceptibility testing methods in a multicenter evaluation of MicroScan dried overnight gram-positive MIC panels for detection of vancomycin and high-level aminoglycoside resistances in enterococci. J. Clin. Microbiol. 36:2996-3001. - PMC - PubMed
-
- Chow, J. W. 2000. Aminoglycoside resistance in enterococci. Clin. Infect. Dis. 31:586-589. - PubMed
-
- d'Azevedo, P. A., C. A. Dias, A. L. S. Gonçalves, F. Rowe, and L. M. Teixeira. 2001. Evaluation of an automated system for the identification and antimicrobial susceptibility testing of enterococci. Diagn. Microbiol. Infect. Dis. 40:157-161. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Molecular Biology Databases