The clinical relevance of electromyopotential oversensing in current unipolar devices
- PMID: 1279592
- DOI: 10.1111/j.1540-8159.1992.tb03014.x
The clinical relevance of electromyopotential oversensing in current unipolar devices
Abstract
Electromyopotential oversensing of unipolar pacemakers was first appreciated 20 years ago, but its prevalence in present day devices is less well defined. Thirty-four pacemaker patients, only two with symptoms suggestive of oversensing, were evaluated in unipolar settings to assess the frequency of provocation of oversensing in one or, if present, both pacing channels. The sensing threshold of each patient, whenever possible, was recorded as well.
Results: Atrial oversensing occurred in 11/18 patients (61%), all at sensitivities in the 0.4-1.0 mV range. Ventricular oversensing was noted in 13/33 patients (39%), with all but one programmed to settings of 1.25 mV or more sensitive (i.e. < 1.25 mV). Twenty six of 26 patients amenable to testing had ventricular sensing thresholds of at least 4.0 mV or more. Of the 15 patients amenable to atrial sensing threshold testing, 4 had a threshold of 1.0 mV or < 1.0 mV, 6 had thresholds between 1.0-2.0 mV, and 5 sensed at settings > 2.0 mV.
Conclusion: Electromyopotential oversensing remains a relevant issue in current day unipolar pacemakers. Most patients do not describe symptoms related to electromyopotential interference, yet such interference is frequently provoked. Oversensing is common at high sensitivities typically utilized for atrial sensing, but quite unusual at settings necessary for adequate ventricular sensing. Programming unipolar devices to unnecessarily high sensitivities should be avoided or serious consequences may result.
Similar articles
-
Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.J Interv Card Electrophysiol. 1998 Mar;2(1):33-40. doi: 10.1023/a:1009708606035. J Interv Card Electrophysiol. 1998. PMID: 9869994 Clinical Trial.
-
Should unipolar leads be implanted in the atrium? A Holter electrocardiographic comparison of threshold adapted unipolar and high sensitive bipolar sensing.Pacing Clin Electrophysiol. 1998 Aug;21(8):1601-8. doi: 10.1111/j.1540-8159.1998.tb00249.x. Pacing Clin Electrophysiol. 1998. PMID: 9725160
-
VDD(R) pacing: short- and long-term stability of atrial sensing with a single lead system.Pacing Clin Electrophysiol. 1996 Apr;19(4 Pt 1):455-64. doi: 10.1111/j.1540-8159.1996.tb06516.x. Pacing Clin Electrophysiol. 1996. PMID: 8848393
-
Efficacy of prophylactic epicardial pacing leads in children and young adults.Ann Thorac Surg. 2004 Jul;78(1):197-202; discussion 202-3. doi: 10.1016/j.athoracsur.2004.02.008. Ann Thorac Surg. 2004. PMID: 15223427 Review.
-
Update on cardiac pacemakers: description, complications, indications, and followup.Cardiovasc Clin. 1985;16(1):177-213. Cardiovasc Clin. 1985. PMID: 3915711 Review.
Cited by
-
[Guidelines for heart pacemaker therapy].Z Kardiol. 2005 Oct;94(10):704-20. doi: 10.1007/s00392-005-0269-3. Z Kardiol. 2005. PMID: 16200488 German. No abstract available.
-
Unipolar sensing in contemporary pacemakers: using myopotential testing to define optimal sensitivity settings.J Interv Card Electrophysiol. 1998 Mar;2(1):33-40. doi: 10.1023/a:1009708606035. J Interv Card Electrophysiol. 1998. PMID: 9869994 Clinical Trial.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical