Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries
- PMID: 12799888
- DOI: 10.1007/s00464-002-8833-7
Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis of small-, medium-, and large-sized arteries
Abstract
Background: Advanced laparoscopic procedures have necessitated the development of new technology for vascular control. Suture ligation can be time-consuming and cumbersome during laparoscopic dissection. Titanium clips have been used for hemostasis, and recently plastic clips and energy sources such as ultrasonic coagulating shears and bipolar thermal energy devices have become popular. The purpose of this study was to compare the bursting pressure of arteries sealed with ultrasonic coagulating shears (UCS), electrothermal bipolar vessel sealer (EBVS), titanium laparoscopic clips (LCs), and plastic laparoscopic clips (PCs). In addition, the spread of thermal injury from the UCS and the EBVS was compared.
Methods: Arteries in three size groups (2-3, 4-5 and 6-7 mm) were harvested from freshly euthanized pigs. Each of the four devices was used to seal 16 specimens from each size group for burst testing. A 5-Fr catheter was placed into the open end of the specimen and secured with a purse-string suture. The catheter was connected to a pressure monitor and saline was infused until there was leakage from the sealed end. This defined the bursting pressure in mmHg. The ultrasonic shears and bipolar thermal device were used to seal an additional 8 vessels in each size group, which were sent for histologic examination. These were examined with hematoxylin and eosin stains, and the extent of thermal injury, defined by coagulation necrosis, was measured in millimeters. Analysis of variance was performed and, where appropriate, a Tukey's test was also performed.
Results: The EBVS's mean burst pressure was statistically higher than that of the UCS at 4 or 5 mm (601 vs 205 mmHg) and 6 or 7 mm (442 vs 175 mmHg). EBVS had higher burst pressures for the 4 or 5-mm group (601 mmHg) and 6 or 7-mm group (442 mmHg) compared with its pressure at 2 or 3 mm (128 mmHg) ( p = 0.0001). The burst pressures of the UCS and EBVS at 2 or 3 mm were not significantly different. Both clips were statistically stronger than the thermal devices except at 4 or 5 mm, in which case the EBVS was as strong as the LC (601 vs 593 mmHg). The PC and LC were similar except at 4 or 5 mm, where the PC was superior (854 vs 593 mmHg). The PC burst pressure for 4 or 5 mm (854 mmHg) was statistically higher than that for vessels 2 or 3 mm (737 mmHg) but not different from the 6 or 7 mm pressure (767 mmHg). Thermal spread was not statistically different when comparing EBVS and UCS at any size (EBVS mean = 2.57 mm vs UCS mean = 2.18 mm).
Conclusions: Both the PC and LC secured all vessel sizes to well above physiologic levels. The EBVS can be used confidently in vessels up to 7 mm. There is no difference in the thermal spread of the LigaSure vessel sealer and the UCS.
Similar articles
-
Effectiveness of the ultrasonic coagulating shears, LigaSure vessel sealer, and surgical clip application in biliary surgery: a comparative analysis.Am Surg. 2001 Sep;67(9):901-6. Am Surg. 2001. PMID: 11565773
-
Electrothermal bipolar vessel sealing device vs. ultrasonic coagulating shears in laparoscopic colectomies: a comparative study.Surg Endosc. 2007 Sep;21(9):1526-31. doi: 10.1007/s00464-006-9143-2. Epub 2007 Feb 8. Surg Endosc. 2007. PMID: 17287913
-
Comparison of linear and torsional mode ultrasonic coagulating shears for the sealing of medium- to large-sized arteries.Surg Endosc. 2007 Jul;21(7):1165-9. doi: 10.1007/s00464-006-9113-8. Epub 2006 Dec 6. Surg Endosc. 2007. PMID: 17149550
-
A review of currently available vessel sealing systems.Minim Invasive Ther Allied Technol. 2007;16(1):52-7. doi: 10.1080/13645700601181414. Minim Invasive Ther Allied Technol. 2007. PMID: 17365677 Review.
-
Novel hemostatic devices in thyroid surgery: electrothermal bipolar vessel sealing system and harmonic scalpel.Expert Rev Med Devices. 2008 Jul;5(4):447-66. doi: 10.1586/17434440.5.4.447. Expert Rev Med Devices. 2008. PMID: 18573045 Review.
Cited by
-
Chylous ascites after colorectal cancer surgery: risk factors and impact on short-term and long-term outcomes.Langenbecks Arch Surg. 2016 Dec;401(8):1171-1177. doi: 10.1007/s00423-016-1500-6. Epub 2016 Aug 24. Langenbecks Arch Surg. 2016. PMID: 27553111
-
The WoundWand and Its Novel Use in Burn Excision Surgery.Eplasty. 2015 Jan 8;15:e1. eCollection 2015. Eplasty. 2015. PMID: 25671049 Free PMC article.
-
Thermal fusion: effects and interactions of temperature, compression, and duration variables.Surg Endosc. 2012 Dec;26(12):3626-33. doi: 10.1007/s00464-012-2386-1. Epub 2012 Jul 7. Surg Endosc. 2012. PMID: 22773229
-
Sympathetic chain clipping for hyperhidrosis is not a reversible procedure.Surg Endosc. 2013 Aug;27(8):3043. doi: 10.1007/s00464-013-2815-9. Epub 2013 Feb 8. Surg Endosc. 2013. PMID: 23392986 No abstract available.
-
Bipolar Vessel Sealing System Versus Suture Ligation in Selective Neck Dissection.J Maxillofac Oral Surg. 2018 Dec;17(4):495-501. doi: 10.1007/s12663-017-1064-1. Epub 2017 Dec 11. J Maxillofac Oral Surg. 2018. PMID: 30344392 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical