Correction of compliance and resistance altered by endotracheal tube leaks
- PMID: 12831418
- DOI: 10.1097/01.PCC.0000059429.95034.73
Correction of compliance and resistance altered by endotracheal tube leaks
Abstract
Objective: Measurements of lung compliance and resistance are influenced by endotracheal tube leaks. To keep compliance and resistance reliable, we developed an algorithm to correct inspiratory and expiratory volume and flow mathematically.
Design: Prospective, clinical study.
Setting: University research laboratory and neonatal intensive care unit.
Model: A ventilated lung model with a linear pressure-volume relationship and with adjustment of an increasing endotracheal tube leak was investigated.
Patients: A total of 21 ventilated premature neonates (median weight, 1220 g; range, 640-2160 g; median leak, 32%; range, 24-56%) were studied.
Measurements and main results: Compliance and resistance were calculated from the recordings of flow, volume, and airway pressure over time employing linear regression of the equation of motion to obtain compliance and resistance. Compliance and resistance altered by leaks were corrected and compared with measurements without leak. Compliance and resistance of the lung model could be corrected up to an endotracheal tube leak size of 86%. Compliance and resistance without leak and after leak correction did not differ significantly for all infants using the linear regression method (p >.05). For the correction of compliance in 15 and for the correction of resistance in 12 of the 21 infants, the coefficients of variation of ten measured breaths without leak were greater or equal to the differences of the values of compliance and resistance between conditions of no leak and corrected leak, respectively.
Conclusion: Pulmonary compliance and resistance can be reliably corrected even in the presence of a substantial endotracheal tube leak, which makes pulmonary function tests more reliable.
Similar articles
-
Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships.Eur J Appl Physiol. 2007 Feb;99(3):301-11. doi: 10.1007/s00421-006-0339-x. Epub 2007 Jan 6. Eur J Appl Physiol. 2007. PMID: 17206444
-
Calculation of intratracheal airway pressure in ventilated neonatal piglets with endotracheal tube leaks.Crit Care Med. 2007 May;35(5):1383-9. doi: 10.1097/01.CCM.0000260244.32756.BB. Crit Care Med. 2007. PMID: 17414085
-
Limitation of measurements of expiratory tidal volume and expiratory compliance under conditions of endotracheal tube leaks.Pediatr Crit Care Med. 2008 Jan;9(1):69-75. doi: 10.1097/01.PCC.0000298660.16328.BA. Pediatr Crit Care Med. 2008. PMID: 18477916
-
Lung function tests in neonates and infants with chronic lung disease: forced expiratory maneuvers.Pediatr Pulmonol. 2006 Mar;41(3):199-214. doi: 10.1002/ppul.20320. Pediatr Pulmonol. 2006. PMID: 16288484 Review.
-
Lung function tests in neonates and infants with chronic lung disease of infancy: functional residual capacity.Pediatr Pulmonol. 2006 Jan;41(1):1-22. doi: 10.1002/ppul.20318. Pediatr Pulmonol. 2006. PMID: 16331641 Review.
Cited by
-
Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships.Eur J Appl Physiol. 2007 Feb;99(3):301-11. doi: 10.1007/s00421-006-0339-x. Epub 2007 Jan 6. Eur J Appl Physiol. 2007. PMID: 17206444
-
Effect of the Y-piece of the ventilation circuit on ventilation requirements in extremely low birth weight infants.Intensive Care Med. 2005 Aug;31(8):1095-100. doi: 10.1007/s00134-005-2711-4. Epub 2005 Jul 6. Intensive Care Med. 2005. PMID: 15999252
-
Ventilator assessment of respiratory mechanics in paediatric intensive care.Eur J Pediatr. 2008 Mar;167(3):287-91. doi: 10.1007/s00431-007-0477-0. Epub 2007 Mar 30. Eur J Pediatr. 2008. PMID: 17394017 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical