The PEP respiratory monitor: a validation study
- PMID: 12835340
- PMCID: PMC1726120
- DOI: 10.1136/emj.20.4.326
The PEP respiratory monitor: a validation study
Abstract
The search for a reliable and accurate respiratory rate monitor for use in non-intubated patients has proved to be a long and fruitless one. A new device fulfilling the criteria for such a monitor has recently been described. The pyroelectric polymer (PEP) device is safe, non-invasive, and cheap. In this study the PEP device, transthoracic impedance, and standard observer counting were all compared with the existing gold standard of capnography in 12 healthy adult volunteers. Using a standard statistical technique it was shown that the PEP device performed as well as a capnograph and was more accurate than the other currently available methods of monitoring respiratory rate.