Ultrasound biomicroscopy and Scheimpflug photography of angle-supported phakic intraocular lens for high myopia
- PMID: 12842684
- DOI: 10.1016/s0886-3350(02)02045-x
Ultrasound biomicroscopy and Scheimpflug photography of angle-supported phakic intraocular lens for high myopia
Abstract
Purpose: To evaluate the equivalence of Scheimpflug photography (SP) and ultrasound biomicroscopy (UBM) in determining corneal epithelium-intraocular lens (IOL) and border IOL-iris distances.
Setting: Universität Erlangen-Nuremberg, Erlangen, Germany.
Methods: In 26 eyes of 17 patients who had a NuVita MA20 angle-supported anterior chamber intraocular lens (Chiron-Domilens), SP and UBM were used to evaluate the distance between the endothelium and the anterior IOL face in central and peripheral regions (12 o'clock and 6 o'clock positions) and between the border of the anterior IOL face and the iris. The Wilcoxon test was used for statistical analysis.
Results: The mean central endothelium-anterior IOL face distance was 2.01 mm and 2.00 mm by SP and UBM, respectively. The mean peripheral endothelium-anterior IOL border distance was 1.28 mm and 1.58 mm, respectively, and the mean peripheral anterior IOL face-iris distance, 0.89 mm and 0.75 mm, respectively.
Conclusions: The difference between the 2 methods in the central endothelium-anterior IOL face distance was not significant (methods were equivalent), but the difference in the peripheral endothelium-anterior IOL face distance was. This may be the result of difficulty in obtaining the exact transition point between the IOL and the haptics by SP examination. The difference between the 2 methods in the IOL border-iris distance was also significant because of the irregularity of the iris surface; therefore, measurements were performed at different sites along this structure. The significant differences in the peripheral endothelium-IOL and IOL border-iris distances indicate that although both methods are useful, they are not equivalent.
Similar articles
-
Position of angle-supported, iris-fixated, and ciliary sulcus-implanted myopic phakic intraocular lenses evaluated by Scheimpflug photography.Am J Ophthalmol. 2004 Nov;138(5):723-31. doi: 10.1016/j.ajo.2004.06.009. Am J Ophthalmol. 2004. PMID: 15531305
-
Scheimpflug biometry of the anterior segment after implantation of foldable iris-fixated lenses.J Refract Surg. 2006 Mar;22(3):243-6. doi: 10.3928/1081-597X-20060301-08. J Refract Surg. 2006. PMID: 16602312
-
Three-year stability of an angle-supported foldable hydrophobic acrylic phakic intraocular lens evaluated by Scheimpflug photography.J Cataract Refract Surg. 2010 Jul;36(7):1120-6. doi: 10.1016/j.jcrs.2010.01.023. J Cataract Refract Surg. 2010. PMID: 20610089 Clinical Trial.
-
[Scheimpflug photography for the examination of phakic intraocular lenses].Ophthalmologe. 2014 Oct;111(10):935-41. doi: 10.1007/s00347-013-2963-2. Ophthalmologe. 2014. PMID: 25332044 Review. German.
-
Ultrasound biomicroscopy in the management of complex cataract and intraocular lens: A review.Clin Exp Ophthalmol. 2024 Mar;52(2):186-206. doi: 10.1111/ceo.14321. Epub 2024 Jan 12. Clin Exp Ophthalmol. 2024. PMID: 38214059 Review.
Cited by
-
Differences between Scheimpflug and optical coherence tomography in determining safety distances in eyes with an iris-fixating phakic intraocular lens.Graefes Arch Clin Exp Ophthalmol. 2021 Jan;259(1):231-238. doi: 10.1007/s00417-020-04874-7. Epub 2020 Aug 6. Graefes Arch Clin Exp Ophthalmol. 2021. PMID: 32766949 Free PMC article.
-
Ultrasonography of the Anterior Segment.Ultrasound Clin. 2008 Apr 1;3(2):201-206. doi: 10.1016/j.cult.2008.04.004. Ultrasound Clin. 2008. PMID: 20577648 Free PMC article. No abstract available.
-
[Intraocular lenses for the correction of refraction errors. Part 1: phakic anterior chamber lenses].Ophthalmologe. 2005 Oct;102(10):1003-7; quiz 1018. doi: 10.1007/s00347-005-1271-x. Ophthalmologe. 2005. PMID: 16172789 Review. German.
-
Diagnostic Techniques to Increase the Safety of Phakic Intraocular Lenses.Diagnostics (Basel). 2023 Jul 27;13(15):2503. doi: 10.3390/diagnostics13152503. Diagnostics (Basel). 2023. PMID: 37568866 Free PMC article. Review.
-
[Intraocular lenses for the correction of refraction errors. Part II. Phakic posterior chamber lenses and refractive lens exchange with posterior chamber lens implantation].Ophthalmologe. 2005 Nov;102(11):1105-17; quiz 1118-9. doi: 10.1007/s00347-005-1274-7. Ophthalmologe. 2005. PMID: 16220314 Review. German.