[Comparative study of antibody identification in the gel centrifugation test (ID Microtyping System), solid phase antiglobulin test (Solidscreen Capture R, Ready ID) and tube test]
- PMID: 1284737
[Comparative study of antibody identification in the gel centrifugation test (ID Microtyping System), solid phase antiglobulin test (Solidscreen Capture R, Ready ID) and tube test]
Abstract
The gel centrifugation test (ID Microtyping System)--a new method for antibody screening--was compared with solid phase systems (Solidscreen, Capture R, Ready ID) and the conventional tube test. 141 different antibodies were tested and the results were compared. The ID Microtyping System identified 138 (98%) of all antibodies, the tube test 110 (78%), Capture R 86 (61%), Ready ID 79 (56%) and Solidscreen 75 (53%). The results in identification of all antibodies except cold agglutinins (n = 107) were: in the ID Microtyping System 98% (105), tube test 76% (82), Capture R 70% (75), Ready ID 68% (73) and Solidscreen 61% (65).
Similar articles
-
[Positive direct antiglobulin test (DAGT) using the gel system].Beitr Infusionsther. 1992;30:378-81. Beitr Infusionsther. 1992. PMID: 1284739 German.
-
[Solid phase technique versus gel centrifugation for detection of erythrocyte antibodies--a prospective study comparison the 2 antibody detection tests].Beitr Infusionsther Transfusionsmed. 1996;33:17-21. Beitr Infusionsther Transfusionsmed. 1996. PMID: 8974691 German.
-
[The gel centrifugation test: comparison of antibody identification at 37 degrees C with antibody identification at room temperature].Infusionsther Transfusionsmed. 1993 Jun;20 Suppl 2:61-3. Infusionsther Transfusionsmed. 1993. PMID: 8374292 German.
-
[Solid phase techniques in blood group serology].Infusionsther Transfusionsmed. 1993 Jun;20 Suppl 2:47-51. Infusionsther Transfusionsmed. 1993. PMID: 8374289 Review. German.
-
[Anti-D human immunoglobulin test].Nihon Rinsho. 2005 Jul;63 Suppl 7:719-21. Nihon Rinsho. 2005. PMID: 16111379 Review. Japanese. No abstract available.