Effect of helium-oxygen (heliox) gas mixtures on the function of four pediatric ventilators

Affiliations
Comparative Study

Effect of helium-oxygen (heliox) gas mixtures on the function of four pediatric ventilators

John W Berkenbosch et al. Crit Care Med. 2003 Jul.

Abstract

Objective: To evaluate the effects of helium on the function of four ventilators commonly used in pediatrics: the Bird VIP, Bird VIP Gold, Servo 300, and Servo 900C.

Design: Prospective setting.

Setting: Research laboratory at a university hospital.

Subjects: Helium was administered as an 80:20 mixture of helium-oxygen through the air inlet of the ventilator. Delivered fraction of inspired oxygen (Fio(2)) was compared with the Fio(2) set on the blender dial. Inspiratory displayed tidal volume was recorded as an indicator of what the ventilator "believed" it had delivered and was compared with the V(T) displayed during ventilation with 100% oxygen (control). Actual delivered V(T) was measured by a Neonatal Bicore connected to the side port of a "bag-in-box" spirometer, making measurements independent of inspired gas properties, and was compared with V(T) delivered during ventilation with 100% oxygen.

Interventions: Five gas mixtures were evaluated: Fio(2) = 0.2, 0.4, 0.6, 0.8, and 1.0 (balance helium).

Measurements and main results: Delivered Fio(2) was less than set Fio(2) on the Servo 900C and VIP ventilators. V(T) displayed was minimally altered by helium during volume-controlled ventilation but substantially decreased during pressure-controlled ventilation, particularly with the Bird ventilators. During volume-controlled ventilation, V(T) delivered was substantially increased by helium with the Bird and, to a lesser degree, the Servo 900C ventilators. In contrast, V(T) delivered decreased slightly in helium with the Servo 300. The same pattern, but with a decreased magnitude, was observed for V(T) delivered during pressure-controlled ventilation.

Conclusions: The addition of helium has a significant effect on Fio(2) delivery, displayed inspiratory V(T), and actual delivered V(T) during both volume- and pressure-controlled ventilation in four ventilators commonly used in pediatric critical care. These effects are both ventilator specific and ventilation mode specific, mandating vigilance during helium ventilation in clinical practice.