[The efficacy of carbon-fiber resistive-heating in prevention of core hypothermia during major abdominal surgery]
- PMID: 12854480
[The efficacy of carbon-fiber resistive-heating in prevention of core hypothermia during major abdominal surgery]
Abstract
Background: Perioperative hypothermia causes numerous severe complications, such as coagulopathy, surgical wound infections, and morbid myocardial outcomes. For prevention of intraoperative hypothermia, an inexpensive, non-disposable carbon fiber resistive warming system has been developed.
Methods: We evaluated the efficacy of resistive-heating, comparing to circulating-water mattress and forced-air warming system. Twenty four patients undergoing elective abdominal surgery were randomly assigned to warming with: 1) a circulating water mattress, 2) a lower-body forced-air system, or 3) a carbon-fiber, resistive-heating blanket.
Results: Tympanic membrane temperature in the first two hours of surgery decreased by 1.9 +/- 0.5 degrees C in the water mattress group, 1.0 +/- 0.6 degree C in the forced-air group, 0.8 +/- 0.2 degree C in the resistive-heating group. The decreases in core temperature by the end of surgery were 2.0 +/- 0.8 degrees C in the water mattress group, 0.6 +/- 1.1 degrees C in the forced-air group, and 0.5 +/- 0.4 degree C in the resistive blanket group, respectively. There was no significant difference in the changes of core temperature between the forced-air group and the resistive-heating group. No side effects related to resistive-heating blanket were observed.
Conclusions: Even during major abdominal surgery, carbon-fiber resistive-heating maintains core temperature as effectively as forced air.
Similar articles
-
Resistive-heating and forced-air warming are comparably effective.Anesth Analg. 2003 Jun;96(6):1683-1687. doi: 10.1213/01.ANE.0000062770.73862.B7. Anesth Analg. 2003. PMID: 12760996 Clinical Trial.
-
Core temperatures during major abdominal surgery in patients warmed with new circulating-water garment, forced-air warming, or carbon-fiber resistive-heating system.J Anesth. 2012 Apr;26(2):168-73. doi: 10.1007/s00540-011-1306-1. Epub 2011 Dec 22. J Anesth. 2012. PMID: 22189652 Free PMC article. Clinical Trial.
-
Resistive polymer versus forced-air warming: comparable heat transfer and core rewarming rates in volunteers.Anesth Analg. 2008 Nov;107(5):1621-6. doi: 10.1213/ane.0b013e3181845502. Anesth Analg. 2008. PMID: 18931221
-
Effectiveness of application of carbon-fibre polymer-fabric resistive heating compared with forced-air warming to prevent unintentional intraoperative hypothermia in patients undergoing elective abdominal operations: A systematic review and meta-analysis of randomised controlled trials.J Clin Nurs. 2020 Dec;29(23-24):4429-4439. doi: 10.1111/jocn.15463. Epub 2020 Sep 10. J Clin Nurs. 2020. PMID: 32841437
-
Prevention of perioperative hypothermia with forced-air warming systems and upper-body blankets.Surg Technol Int. 2006;15:19-22. Surg Technol Int. 2006. PMID: 17029156 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical