Corneal astigmatism in applanation tonometry
- PMID: 12855970
- DOI: 10.1038/sj.eye.6700417
Corneal astigmatism in applanation tonometry
Abstract
Purpose: This short study aims at quantitatively correlating scale readings of the Goldmann tonometer to corneal curvatures in one and the same eye.
Methods: A total of 30 eyes of 21 adult patients who presented themselves in clinical practice for routine examination were arbitrarily selected. Corneal curvatures were measured in the horizontal and vertical meridians and then applanation tension measured in the same order. The data were submitted to statistical analysis by computer using Microsoft Excel and SAS System programs. Results All cases exhibited a strong positive correlation of tonometric readings to corneal curvature, that is, the flatter the curvature, the lower the reading on the scale of the tonomer, and vice versa.
Conclusion: The numbers on the scale of the Goldmann tonometer may not represent true intraocular pressure in mmHg because their values depend on a number of variables of the globe, including corneal astigmatism. Attention to these influences may help interpret the tonometric record in difficult cases.
Similar articles
-
Comparison of dynamic contour tonometry with goldmann applanation tonometry.Invest Ophthalmol Vis Sci. 2004 Sep;45(9):3118-21. doi: 10.1167/iovs.04-0018. Invest Ophthalmol Vis Sci. 2004. PMID: 15326129 Clinical Trial.
-
Effects of corneal thickness, curvature, astigmatism and direction of gaze on Goldmann applanation tonometry readings.Ophthalmic Res. 2006;38(1):49-55. doi: 10.1159/000089762. Epub 2005 Nov 16. Ophthalmic Res. 2006. PMID: 16293977
-
Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties.Am J Ophthalmol. 2008 Feb;145(2):215-221. doi: 10.1016/j.ajo.2007.10.003. Am J Ophthalmol. 2008. PMID: 18222191
-
[The role of pachymetry in routine glaucoma diagnosis].Klin Monbl Augenheilkd. 2006 Feb;223(2):117-30. doi: 10.1055/s-2005-858724. Klin Monbl Augenheilkd. 2006. PMID: 16485223 Review. German.
-
New ways to measure intraocular pressure.Curr Opin Ophthalmol. 2008 Mar;19(2):122-6. doi: 10.1097/ICU.0b013e3282f391ae. Curr Opin Ophthalmol. 2008. PMID: 18301285 Review.
Cited by
-
Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.Br J Ophthalmol. 2006 Dec;90(12):1444-6. doi: 10.1136/bjo.2006.102970. Br J Ophthalmol. 2006. PMID: 17114589 Free PMC article. Review.
-
Ocular biometric parameters are associated with non-contact tonometry measured intraocular pressure in non-pathologic myopic patients.Int Ophthalmol. 2020 Feb;40(2):431-437. doi: 10.1007/s10792-019-01203-8. Epub 2019 Dec 2. Int Ophthalmol. 2020. PMID: 31792850
-
The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.Int Ophthalmol. 2020 Apr;40(4):849-857. doi: 10.1007/s10792-019-01248-9. Epub 2019 Dec 2. Int Ophthalmol. 2020. PMID: 31792851
-
Comparison of Intraocular Pressure Measurements Using Three Different Methods (Goldmann Applanation Tonometry (GAT), Corvis ST, and iCare) Following Penetrating Keratoplasty.J Clin Med. 2024 Nov 22;13(23):7046. doi: 10.3390/jcm13237046. J Clin Med. 2024. PMID: 39685505 Free PMC article.
-
Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methods.Clin Ophthalmol. 2014 Mar 26;8:605-10. doi: 10.2147/OPTH.S53544. eCollection 2014. Clin Ophthalmol. 2014. PMID: 24707165 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources