Objective: To determine the feasibility, reliability, and validity of using partial coherence interferometry, a noncontact method that detects interference patterns from various layers of the eye, to measure axial length in young children.
Methods: The right eye of 64 subjects (mean age, 8.4 y; age range, 3.4-12.9 y; best-corrected visual acuity >or=20/30) was measured. Subjects fixated monocularly on the collimated light pattern from a laser diode (the alignment beam) and the operator used a video monitor to align the corneal reflection in the optical path. Axial length was measured during an 0.8-second scan using interference patterns from a collimated short coherence superluminescence diode aligned coaxially with the laser diode. Five series of 16 readings each were obtained. The average axial length for each of the 5 series of readings was calculated. Main Outcome Measure Axial length.
Results: Within-subject precision of axial length measurements was high, with an overall SE of measurement of 8 micro m for individual subjects across the 5 sessions (95% confidence interval, +/-16 micro m). Subgroup analysis showed that sex, age, spherical equivalent, and refractive error exerted statistically significant effects on precision, but all of the differences among subgroups were 3 micro m or less and likely to be insignificant clinically. Axial length measured by partial coherence interferometry varied systematically, with factors known to influence eye length (ie, age and refractive error), further validating the measurement method.
Conclusion: The partial coherence interferometry technique provides reproducible, extraordinarily precise eye length measurements in young children and should enable novel approaches to study eye growth and refractive development.
Similar articles
-
Reliability and validity of the partial coherence interferometry for measurement of ocular axial length in children.Eye (Lond). 2006 Sep;20(9):1021-4. doi: 10.1038/sj.eye.6702069. Epub 2005 Aug 12. Eye (Lond). 2006. PMID: 16096655
-
Evaluation of axial length measurement of the eye using partial coherence interferometry and ultrasound in cases of macular disease.Ophthalmology. 2010 Sep;117(9):1750-4. doi: 10.1016/j.ophtha.2010.01.018. Epub 2010 May 14. Ophthalmology. 2010. PMID: 20471687
-
Macular thickness variations with sex, age, and axial length in healthy subjects: a spectral domain-optical coherence tomography study.Invest Ophthalmol Vis Sci. 2010 Aug;51(8):3913-8. doi: 10.1167/iovs.09-4189. Epub 2010 Mar 31. Invest Ophthalmol Vis Sci. 2010. PMID: 20357206
-
Diurnal axial length fluctuations in human eyes.Invest Ophthalmol Vis Sci. 2004 Jan;45(1):63-70. doi: 10.1167/iovs.03-0294. Invest Ophthalmol Vis Sci. 2004. PMID: 14691155
-
Optical coherence biometry.Dev Ophthalmol. 2002;34:119-30. doi: 10.1159/000060791. Dev Ophthalmol. 2002. PMID: 12520608 Review. No abstract available.
Cited by
-
Axial length measurement using partial coherence interferometry in myopic children: repeatability of the measurement and comparison with refractive components.Jpn J Ophthalmol. 2007 Mar-Apr;51(2):105-10. doi: 10.1007/s10384-006-0410-5. Epub 2007 Apr 6. Jpn J Ophthalmol. 2007. PMID: 17401619
-
Partial coherence interferometry versus immersion ultrasonography for axial length measurement in children.J Cataract Refract Surg. 2010 Dec;36(12):2100-4. doi: 10.1016/j.jcrs.2010.07.013. J Cataract Refract Surg. 2010. PMID: 21111313 Free PMC article.
-
Refractive changes after removal of anterior IOLs in temporary piggyback IOL implantation for congenital cataracts.Korean J Ophthalmol. 2013 Apr;27(2):93-7. doi: 10.3341/kjo.2013.27.2.93. Epub 2013 Feb 27. Korean J Ophthalmol. 2013. PMID: 23543846 Free PMC article. Clinical Trial.
-
Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer.Eye Vis (Lond). 2020 Mar 2;7:13. doi: 10.1186/s40662-020-00177-4. eCollection 2020. Eye Vis (Lond). 2020. PMID: 32161772 Free PMC article.
-
In vivo human choroidal thickness measurements: evidence for diurnal fluctuations.Invest Ophthalmol Vis Sci. 2009 Jan;50(1):5-12. doi: 10.1167/iovs.08-1779. Epub 2008 Aug 21. Invest Ophthalmol Vis Sci. 2009. PMID: 18719079 Free PMC article.
Publication types