Effect of a tight necktie on intraocular pressure
- PMID: 12881330
- PMCID: PMC1771792
- DOI: 10.1136/bjo.87.8.946
Effect of a tight necktie on intraocular pressure
Abstract
Aim: To evaluate the effect of a tight necktie on intraocular pressure (IOP) measurement using Goldmann applanation tonometry.
Methods: 40 eyes of 20 normal subjects and 20 open angle glaucoma patients (all male) were enrolled. IOP was measured with an open shirt collar, 3 minutes after placing a tight necktie, and 3 minutes after loosening it. All measurements were made by the same examiner.
Results: Mean IOP in normal subjects increased by 2.6 (SD 3.9) mm Hg (p=0.008, paired t test; range -3 to +14 mm Hg) and in glaucoma patients by 1.0 (1.8) mm Hg (p=0.02, paired t test; range -2 to +4.5 mm Hg). In normal subjects, IOP in 12 eyes was increased by >/=2 mm Hg and in seven eyes by >/=4 mm Hg. In glaucoma patients, IOP in six eyes was increased by >/=2 mm Hg and in two eyes by >/=4 mm Hg.
Conclusion: A tight necktie increases IOP in both normal subjects and glaucoma patients and could affect the diagnosis and management of glaucoma.
Similar articles
-
Does extended wear of a tight necktie cause raised intraocular pressure?J Glaucoma. 2005 Dec;14(6):508-10. doi: 10.1097/01.ijg.0000185435.08051.cb. J Glaucoma. 2005. PMID: 16276286
-
[Effect of a tight necktie on intraocular pressure].Med Pregl. 2012 Jan-Feb;65(1-2):13-7. doi: 10.2298/mpns1202013b. Med Pregl. 2012. PMID: 22452233 Clinical Trial. Serbian.
-
Impact factors on intraocular pressure measurements in healthy subjects.Br J Ophthalmol. 2004 Dec;88(12):1510-1. doi: 10.1136/bjo.2004.049924. Br J Ophthalmol. 2004. PMID: 15548801 Free PMC article.
-
Attempted eyelid closure affects intraocular pressure measurement in open-angle glaucoma patients.Am J Ophthalmol. 2002 Aug;134(2):186-9. doi: 10.1016/s0002-9394(02)01513-1. Am J Ophthalmol. 2002. PMID: 12140024 Clinical Trial.
-
Clinical comparison of pascal dynamic contour tonometry and goldmann applanation tonometry in asymmetric open-angle glaucoma.J Glaucoma. 2007 Dec;16(8):694-9. doi: 10.1097/IJG.0b013e3180408dc6. J Glaucoma. 2007. PMID: 18091457
Cited by
-
Tight necktie, intraocular pressure, and intracranial pressure.Br J Ophthalmol. 2005 Jun;89(6):786-7; author reply 787. doi: 10.1136/bjo.2005.068544/10.1136/bjo.2005.069757. Br J Ophthalmol. 2005. PMID: 15923535 Free PMC article. No abstract available.
-
Lifestyle, nutrition, and glaucoma.J Glaucoma. 2009 Aug;18(6):423-8. doi: 10.1097/IJG.0b013e31818d3899. J Glaucoma. 2009. PMID: 19680048 Free PMC article. Review.
-
[Assessment of risk factors for the occurrence of open angle glaucoma : Guidelines of the German Ophthalmological Society and the Professional Association of Ophthalmologists in Germany].Ophthalmologe. 2021 Jan;118(Suppl 1):1-14. doi: 10.1007/s00347-020-01169-4. Ophthalmologe. 2021. PMID: 32705323 Review. German. No abstract available.
-
Glaucoma suspects: A practical approach.Taiwan J Ophthalmol. 2018 Apr-Jun;8(2):74-81. doi: 10.4103/tjo.tjo_106_17. Taiwan J Ophthalmol. 2018. PMID: 30038885 Free PMC article. Review.
-
Influence of carbon dioxide insufflation of the neck on intraocular pressure during robot-assisted endoscopic thyroidectomy: a comparison with open thyroidectomy.Surg Endosc. 2013 May;27(5):1587-93. doi: 10.1007/s00464-012-2633-5. Epub 2012 Oct 17. Surg Endosc. 2013. PMID: 23073689
References
-
- Wilson MR, Martone JF. Epidemiology of chronic open-angle glaucoma. In: Ritch R, Shields MB, Krupin T, eds. The glaucomas. Vol 2. 2nd ed. St Louis: CV Mosby, 1996:351–67.
-
- Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 1993;38:1–30. - PubMed
-
- Schottenstein EM. Intraocular pressure and tonometry. In: Ritch R, Shields MB, Krupin T, eds. The glaucomas. Vol 2. 2nd ed. St Louis: CV Mosby, 1996:407–28.
-
- Mark HH. Corneal curvature in applanation tonometry. Am J Ophthalmol 1973;76:223–4. - PubMed
-
- Doughty MJ. Human corneal thickness and its impact on intraocular pressure measures: a review. Surv Ophthalmol 2000;44:367–408. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources