[Evidence of lacrimal plugs via high resolution ultrasound]
- PMID: 12886505
- DOI: 10.1055/s-2003-40944
[Evidence of lacrimal plugs via high resolution ultrasound]
Abstract
Background: The practical value of high-frequency ultrasound (transducer frequency of 20 MHz) for studying lacrimal plugs positioned into canaliculi was proved.
Materials and methods: Twelve patients with twenty intracanalicular plugs and two punctum plugs were examined via high-frequency B-scan ultrasonography using 20 MHz transducer (model I3 Sacramento, USA). Detection and localisation of the intracanalicular plugs was made by a 20 MHz sector scanner. The ultrasound examinations were performed 1 - 24 month after the placement of lacrimal plugs. After patient's head positioning, the high-frequency ultrasound investigation was done via immersion fluid (2 % methylcellulose).
Results: All patients with dry eye treated by lacrimal plug implant showed echographic structure in the lacrimal canaliculus. In transversal echograms it was possible to image both canaliculi together when the lids were half-closed. Contrary to the normal state, it was not necessary to inject viscous fluid into the canaliculus. High-resolution ultrasound was able to differentiate the normal canaliculus from the findings after plug placement. The echograms can vary from one plug type to another. Highly reflective structures were found after the placement of silicone intracanalicular plugs, e. g. HERRICK-Plug. In contrast, the ultrasonic image taken through acrylic polymer intracanalicular plugs showed homogeneous small reflective inner structure, e. g. SMART-Plug. However, smooth and flat acoustic interface between acrylic polymer plug and the lacrimal canaliculus produced strong echoes.
Conclusions: 20 MHz ultrasound seems to be well suited for the detection and localisation of intracanalicular plugs. By use of 20 MHz ultrasound scans it is possible to get high-quality images of the intracanalicular plug and around lacrimal canaliculus. Compared with UBM, the depth of penetration is much higher with negligible resolution. On the whole, we believe that 20 MHz ultrasound can become a useful tool for evaluating the placement of intracanalicular plugs after insertion.
Comment in
-
["High-tech" in diagnostic procedures of the lacrimal system].Klin Monbl Augenheilkd. 2003 Jul;220(7):459. doi: 10.1055/s-2003-40945. Klin Monbl Augenheilkd. 2003. PMID: 12886504 German. No abstract available.
Similar articles
-
20-MHz ultrasound and its value in imaging of lacrimal plugs.Ophthalmologica. 2004 Jan-Feb;218(1):14-9. doi: 10.1159/000074561. Ophthalmologica. 2004. PMID: 14688430
-
[20-MHz ultrasound of pre-saccular lacrimal ducts].Ophthalmologe. 2002 Jan;99(1):25-8. Ophthalmologe. 2002. PMID: 11840791 German.
-
Identification of retained intracanalicular plugs with ultrasound biomicroscopy.Can J Ophthalmol. 2004 Aug;39(5):533-7. doi: 10.1016/s0008-4182(04)80144-3. Can J Ophthalmol. 2004. PMID: 15491039
-
Safety and Efficacy of Lacrimal Drainage System Plugs for Dry Eye Syndrome: A Report by the American Academy of Ophthalmology.Ophthalmology. 2015 Aug;122(8):1681-7. doi: 10.1016/j.ophtha.2015.04.034. Epub 2015 May 30. Ophthalmology. 2015. PMID: 26038339 Review.
-
Punctal and canalicular plugs: Indications, efficacy and safety.J Fr Ophtalmol. 2019 Mar;42(3):e95-e104. doi: 10.1016/j.jfo.2018.12.003. Epub 2019 Jan 26. J Fr Ophtalmol. 2019. PMID: 30692031 Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources