Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation
- PMID: 12889721
- PMCID: PMC227048
Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation
Abstract
The objective of this prospective clinical study was to evaluate the accuracy of pulse oximetry and capnography in healthy and compromised horses during general anesthesia with spontaneous and controlled ventilation. Horses anesthetized in a dorsal recumbency position for arthroscopy (n = 20) or colic surgery (n = 16) were instrumented with an earlobe probe from the pulse oximeter positioned on the tip of the tongue and a sample line inserted at the Y-piece for capnography. The horses were allowed to breathe spontaneously (SV) for the first 20 min after induction, and thereafter ventilation was controlled (IPPV). Arterial blood, for blood gas analysis, was drawn 20 min after induction and 20 min after IPPV was started. Relationships between oxygen saturation as determined by pulse oximetry (SpO2), arterial oxygen saturation (SaO2), arterial carbon dioxide partial pressure (PaCO2), and end tidal carbon dioxide (P(et)CO2), several physiological variables, and the accuracy of pulse oximetry and capnography, were evaluated by Bland-Altman or regression analysis. In the present study, both SpO2 and P(et)CO2 provided a relatively poor indication of SaO2 and PaCO2, respectively, in both healthy and compromised horses, especially during SV. A difference in heart rate obtained by pulse oximetry, ECG, or palpation is significantly correlated with any pulse oximeter inaccuracy. If blood gas analysis is not available, ventilation to P(et)CO2 of 35 to 45 mmHg should maintain the PaCO2 within a normal range. However, especially in compromised horses, it should never substitute blood gas analysis.
Figures


Similar articles
-
Accuracy of a reflectance pulse oximeter in anesthetized horses.Am J Vet Res. 1993 Apr;54(4):497-501. Am J Vet Res. 1993. PMID: 8484568
-
Evaluation of pulse oximetry in anaesthetised foals using multiple combinations of transducer type and transducer attachment site.Equine Vet J. 1996 Nov;28(6):437-45. doi: 10.1111/j.2042-3306.1996.tb01615.x. Equine Vet J. 1996. PMID: 9049492
-
Evaluation of pulse oximetry in horses surgically treated for colic.Equine Vet J. 1994 Mar;26(2):114-6. doi: 10.1111/j.2042-3306.1994.tb04347.x. Equine Vet J. 1994. PMID: 8575371
-
Respiratory monitoring: arterial blood gas analysis, pulse oximetry, and end-tidal carbon dioxide analysis.Clin Tech Small Anim Pract. 1999 Nov;14(4):227-30. doi: 10.1016/S1096-2867(99)80015-2. Clin Tech Small Anim Pract. 1999. PMID: 10652840 Review.
-
[Non-invasive patient monitoring in veterinary medicine: pulse oximetry and capnography. II. Capnography].Tierarztl Prax. 1995 Feb;23(1):1-16. Tierarztl Prax. 1995. PMID: 7792768 Review. German.
Cited by
-
Running a genetic stop sign accelerates oxygen metabolism and energy production in horses.Science. 2025 Mar 28;387(6741):eadr8589. doi: 10.1126/science.adr8589. Epub 2025 Mar 28. Science. 2025. PMID: 40146832
-
Use of Electrical Impedance Tomography (EIT) to Estimate Tidal Volume in Anaesthetized Horses Undergoing Elective Surgery.Animals (Basel). 2021 May 10;11(5):1350. doi: 10.3390/ani11051350. Animals (Basel). 2021. PMID: 34068514 Free PMC article.
-
Causes, Effects and Methods of Monitoring Gas Exchange Disturbances during Equine General Anaesthesia.Animals (Basel). 2021 Jul 9;11(7):2049. doi: 10.3390/ani11072049. Animals (Basel). 2021. PMID: 34359177 Free PMC article. Review.
-
The role of pulse oximetry in chiropractic practice: a rationale for its use.J Chiropr Med. 2012 Jun;11(2):127-33. doi: 10.1016/j.jcm.2011.10.004. J Chiropr Med. 2012. PMID: 23204957 Free PMC article.
-
Pulse oximetry at two sensor placement sites in conscious foals.Acta Vet Scand. 2025 Jan 23;67(1):6. doi: 10.1186/s13028-025-00794-w. Acta Vet Scand. 2025. PMID: 39849542 Free PMC article.
References
-
- Gillespie JR, Tyler WS, Hall LW. Cardiopulmonary dysfunction in anaesthetised laterally recumbent horses. Am J Vet Res 1969;30:61–72. - PubMed
-
- Meyer RE, Short CE. Arterial to end-tidal CO2 tension and alveolar dead space in halothane — or isoflurane — anesthetized ponies. Am J Vet Res 1985;46:597–599. - PubMed
-
- McDonell W. Respiratory System. In: Thurmon JC, Tranquili WJ, Benson JG, eds. Lumb and Jones' Veterinary Anesthesia. 3rd ed., Wiliams & Wilkins, 1996;115–147.
-
- Moens Y. Arterio-alveolar carbon-dioxide tension difference and alveolar dead space in halothane anaesthetized horses. Equine Vet J 1989;6:282–284. - PubMed
-
- Nyman G, Hedenstierna G. Ventilation-perfusion relationships in the anaesthetized horse. Equine Vet J 1989;21:274–281. - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources