. 2003 Jul;67(3):169-74.

Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation

Affiliations

Accuracy of pulse oximetry and capnography in healthy and compromised horses during spontaneous and controlled ventilation

Judith Koenig et al. Can J Vet Res. 2003 Jul.

Abstract

The objective of this prospective clinical study was to evaluate the accuracy of pulse oximetry and capnography in healthy and compromised horses during general anesthesia with spontaneous and controlled ventilation. Horses anesthetized in a dorsal recumbency position for arthroscopy (n = 20) or colic surgery (n = 16) were instrumented with an earlobe probe from the pulse oximeter positioned on the tip of the tongue and a sample line inserted at the Y-piece for capnography. The horses were allowed to breathe spontaneously (SV) for the first 20 min after induction, and thereafter ventilation was controlled (IPPV). Arterial blood, for blood gas analysis, was drawn 20 min after induction and 20 min after IPPV was started. Relationships between oxygen saturation as determined by pulse oximetry (SpO2), arterial oxygen saturation (SaO2), arterial carbon dioxide partial pressure (PaCO2), and end tidal carbon dioxide (P(et)CO2), several physiological variables, and the accuracy of pulse oximetry and capnography, were evaluated by Bland-Altman or regression analysis. In the present study, both SpO2 and P(et)CO2 provided a relatively poor indication of SaO2 and PaCO2, respectively, in both healthy and compromised horses, especially during SV. A difference in heart rate obtained by pulse oximetry, ECG, or palpation is significantly correlated with any pulse oximeter inaccuracy. If blood gas analysis is not available, ventilation to P(et)CO2 of 35 to 45 mmHg should maintain the PaCO2 within a normal range. However, especially in compromised horses, it should never substitute blood gas analysis.

PubMed Disclaimer

Figures

None
Figure 1. Difference in pulse oximeter oxygen saturation and arterial oxygen saturation readings compared to arterial blood gas values in healthy (arthroscopy group) and compromised (colic group) horses during spontaneous and controlled ventilation. The mean difference (bias) and limits of agreement (± 2 s) are shown.
None
Figure 2. Differences in endtidal CO2 and arterial CO2 readings compared to arterial blood gas values in healthy (arthroscopy group) and compromised (colic group) horses during spontaneous and controlled ventilation. The mean difference (bias) and limits of agreement (± 2 s) are shown.

Similar articles

Cited by

References

    1. Gillespie JR, Tyler WS, Hall LW. Cardiopulmonary dysfunction in anaesthetised laterally recumbent horses. Am J Vet Res 1969;30:61–72. - PubMed
    1. Meyer RE, Short CE. Arterial to end-tidal CO2 tension and alveolar dead space in halothane — or isoflurane — anesthetized ponies. Am J Vet Res 1985;46:597–599. - PubMed
    1. McDonell W. Respiratory System. In: Thurmon JC, Tranquili WJ, Benson JG, eds. Lumb and Jones' Veterinary Anesthesia. 3rd ed., Wiliams & Wilkins, 1996;115–147.
    1. Moens Y. Arterio-alveolar carbon-dioxide tension difference and alveolar dead space in halothane anaesthetized horses. Equine Vet J 1989;6:282–284. - PubMed
    1. Nyman G, Hedenstierna G. Ventilation-perfusion relationships in the anaesthetized horse. Equine Vet J 1989;21:274–281. - PubMed

MeSH terms

LinkOut - more resources