Abstract
Purpose: To determine peak skin dose (PSD), a measure of the likelihood of radiation-induced skin effects, for a variety of common interventional radiology and interventional neuroradiology procedures, and to identify procedures associated with a PSD greater than 2 Gy.
Materials and methods: An observational study was conducted at seven academic medical centers in the United States. Sites prospectively contributed demographic and radiation dose data for subjects undergoing 21 specific procedures in a fluoroscopic suite equipped with built-in dosimetry capability. Comprehensive physics evaluations and periodic consistency checks were performed on each unit to verify the stability and consistency of the dosimeter. Seven of 12 fluoroscopic suites in the study were equipped with skin dose mapping software.
Results: Over a 3-year period, skin dose data were recorded for 800 instances of 21 interventional radiology procedures. Wide variation in PSD was observed for different instances of the same procedure. Some instances of each procedure we studied resulted in a PSD greater than 2 Gy, except for nephrostomy, pulmonary angiography, and inferior vena cava filter placement. Some instances of transjugular intrahepatic portosystemic shunt (TIPS) creation, renal/visceral angioplasty, and angiographic diagnosis and therapy of gastrointestinal hemorrhage produced PSDs greater than 3 Gy. Some instances of hepatic chemoembolization, other tumor embolization, and neuroembolization procedures in the head and spine produced PSDs greater than 5 Gy. In a subset of 709 instances of higher-dose procedures, there was good overall correlation between PSD and cumulative dose (r = 0.86; P <.000001) and between PSD and dose-area-product (r = 0.85, P <.000001), but there was wide variation in these relationships for individual instances.
Conclusions: There are substantial variations in PSD among instances of the same procedure and among different procedure types. Most of the procedures observed may produce a PSD sufficient to cause deterministic effects in skin. It is suggested that dose data be recorded routinely for TIPS creation, angioplasty in the abdomen or pelvis, all embolization procedures, and especially for head and spine embolization procedures. Measurement or estimation of PSD is the best method for determining the likelihood of radiation-induced skin effects. Skin dose mapping is preferable to a single-point measurement of PSD.
Similar articles
-
Radiation doses in interventional radiology procedures: the RAD-IR study: part I: overall measures of dose.J Vasc Interv Radiol. 2003 Jun;14(6):711-27. doi: 10.1097/01.rvi.0000079980.80153.4b. J Vasc Interv Radiol. 2003. PMID: 12817038
-
Contemporary Interventional Radiology Dosimetry: Analysis of 4,784 Discrete Procedures at a Single Institution.J Am Coll Radiol. 2018 Sep;15(9):1214-1221. doi: 10.1016/j.jacr.2018.06.004. Epub 2018 Jul 3. J Am Coll Radiol. 2018. PMID: 29980351
-
Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures.J Vasc Interv Radiol. 2002 Apr;13(4):391-7. doi: 10.1016/s1051-0443(07)61742-4. J Vasc Interv Radiol. 2002. PMID: 11932370
-
Avoidance of radiation injuries from medical interventional procedures.Ann ICRP. 2000;30(2):7-67. doi: 10.1016/S0146-6453(01)00004-5. Ann ICRP. 2000. PMID: 11459599 Review.
-
Radiation dose in interventional fluoroscopic procedures.Appl Radiat Isot. 1999 Jan;50(1):173-84. doi: 10.1016/s0969-8043(98)00046-3. Appl Radiat Isot. 1999. PMID: 10028636 Review.
Cited by
-
Reduction of radiation exposure to operating physician and assistant using a real-time auditory feedback dosimeter during femoral artery puncturing: a study on swine model.Eur Radiol Exp. 2019 Sep 23;3(1):38. doi: 10.1186/s41747-019-0116-3. Eur Radiol Exp. 2019. PMID: 31549259 Free PMC article.
-
Neurointerventions in children: radiation exposure and its import.AJNR Am J Neuroradiol. 2014 Apr;35(4):650-6. doi: 10.3174/ajnr.A3758. Epub 2013 Oct 24. AJNR Am J Neuroradiol. 2014. PMID: 24157736 Free PMC article.
-
AAPM Medical Physics Practice Guideline 12.a: Fluoroscopy dose management.J Appl Clin Med Phys. 2022 Mar;23(3):e13526. doi: 10.1002/acm2.13526. Epub 2022 Feb 17. J Appl Clin Med Phys. 2022. PMID: 35174964 Free PMC article. No abstract available.
-
Percutaneous closure of interatrial communications in adults - prospective embolism prevention study with two- and three-dimensional echocardiography.Cardiovasc Ultrasound. 2004 May 19;2:5. doi: 10.1186/1476-7120-2-5. Cardiovasc Ultrasound. 2004. PMID: 15151699 Free PMC article. Clinical Trial.
-
Radiation risk from fluoroscopically-assisted anterior cruciate ligament reconstruction.Ann R Coll Surg Engl. 2010 May;92(4):330-4. doi: 10.1308/003588410X12628812459814. Ann R Coll Surg Engl. 2010. PMID: 20501019 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources