Elimination of importance factors for clinically accurate selection of beam orientations, beam weights and wedge angles in conformal radiation therapy

Affiliations
Comparative Study

Elimination of importance factors for clinically accurate selection of beam orientations, beam weights and wedge angles in conformal radiation therapy

James L Bedford et al. Med Phys. 2003 Jul.

Abstract

A method of simultaneously optimizing beam orientations, beam weights, and wedge angles for conformal radiotherapy is presented. This method removes the need for importance factors by optimizing one objective only, subject to a set of rigid constraints. This facilitates the production of inverse solutions which, without trial-and-error modification of importance factors, precisely satisfy the specified constraints. The algorithm minimizes an objective function which is based upon the single objective to be optimized, but which is forced to an artificially high value when the constraints are not met, so that only satisfactory solutions are allowed. Due to the complex nature of the objective function space, including multiple local minima separated by large regions of plateau, a random search technique equivalent to fast simulated annealing is used for producing inverse plans. To illustrate the novel features of the new algorithm, a simulation is first presented, for the case of a cylindrical phantom. The morphology of the objective function space is shown to be significantly different for the new algorithm, compared to that for a conventional quadratic objective function. Clinical cases for prostate and craniopharyngioma are then presented. For the prostate case, the objective is to reduce irradiated rectal volume. Three-field, four-field, and six-field optimizations, with or without orientation optimization, are shown to provide solutions which are consistent with previously reported plans and class solutions. For the craniopharyngioma case, which involves the use of a high-precision stereotactic conformal technique, the objective is to reduce the irradiated volume of normal brain. Practically feasible beam angles are produced which, compared to a standard plan, provide a small but worthwhile sparing of normal brain. The algorithm is thereby shown to be robust and suitable for clinical application.