Impact of valve prosthesis-patient mismatch on short-term mortality after aortic valve replacement
Abstract
Background: The prosthesis used for aortic valve replacement (AVR) can be too small in relation to body size, thus causing valve prosthesis-patient mismatch (PPM) and abnormally high transvalvular pressure gradients. This study examined if there is a relation between PPM and short-term mortality after operation.
Methods and results: The indexed valve effective orifice area (EOA) was estimated for each type and size of prosthesis being implanted in 1266 consecutive patients and used to define PPM as not clinically significant if >0.85 cm2/m2, as moderate if >0.65 cm2/m2 and <or=0.85 cm2/m2, and as severe if <or=0.65 cm2/m2; it was correlated with 30-day mortality and compared with other relevant variables. Moderate or severe PPM was present in 38% of patients. Thirty-day mortality was 4.6% (58/1266 patients) and the strongest independent predictors in multivariate analysis were left ventricular ejection fraction <40% (P=0.007), infectious endocarditis (P=0.002), emergent/salvage operation (P=0.002), cardiopulmonary bypass time >120 minutes (P=0.001), and PPM (P=0.003). Relative risk of mortality was increased 2.1-fold (95% confidence interval, 1.2 to 3.7) in patients with moderate PPM and 11.4-fold (4.4 to 29.5) in those with severe PPM. Moreover, risk of mortality for every category of PPM was higher in patients with a left ventricular ejection fraction <40% as compared with >or=40% (nonsignificant PPM, 2.7 versus 1.0; moderate PPM, 7.1 versus 1.8; severe PPM, 77.1 versus 11.3).
Conclusions: PPM is a strong and independent predictor of short-term mortality among patients undergoing AVR, and its impact is related both to its degree of severity and the status of left ventricular function. In contrast to other risk factors, moderate-severe PPM can be largely avoided with the use of a prospective strategy at the time of operation.
Comment in
-
Asymptomatic left ventricular systolic dysfunction--a pending disaster.Circulation. 2003 Aug 26;108(8):e9014-5. doi: 10.1161/01.CIR.0000093160.98631.5D. Circulation. 2003. PMID: 12939252 No abstract available.
Similar articles
-
Impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: influence of age, obesity, and left ventricular dysfunction.J Am Coll Cardiol. 2009 Jan 6;53(1):39-47. doi: 10.1016/j.jacc.2008.09.022. J Am Coll Cardiol. 2009. PMID: 19118723
-
Prosthesis-patient mismatch is associated with higher operative mortality following aortic valve replacement.Heart Lung Circ. 2007 Aug;16(4):260-4. doi: 10.1016/j.hlc.2007.02.082. Epub 2007 Apr 17. Heart Lung Circ. 2007. PMID: 17442619
-
Is patient-prosthesis mismatch an independent risk factor for early and mid-term overall mortality in adult patients undergoing aortic valve replacement?Interact Cardiovasc Thorac Surg. 2009 Sep;9(3):510-8. doi: 10.1510/icvts.2009.207597. Epub 2009 Jun 4. Interact Cardiovasc Thorac Surg. 2009. PMID: 19497953 Review.
-
Effects of patient-prosthesis mismatch on postoperative early mortality in isolated aortic stenosis.J Heart Valve Dis. 2009 Jan;18(1):18-27. J Heart Valve Dis. 2009. PMID: 19301549
-
Patient prosthesis mismatch affects short- and long-term outcomes after aortic valve replacement.Eur J Cardiothorac Surg. 2006 Jul;30(1):15-9. doi: 10.1016/j.ejcts.2006.04.007. Epub 2006 May 26. Eur J Cardiothorac Surg. 2006. PMID: 16730449 Review.
Cited by
-
Patient-Prosthesis Mismatch in Contemporary Small-Size Mechanical Prostheses Does Not Impact Survival at 10 Years.J Cardiovasc Dev Dis. 2022 Jan 31;9(2):48. doi: 10.3390/jcdd9020048. J Cardiovasc Dev Dis. 2022. PMID: 35200701 Free PMC article.
-
Figure of eight suture technique in aortic valve replacement decreases prosthesis-patient mismatch.J Cardiothorac Surg. 2023 Apr 10;18(1):117. doi: 10.1186/s13019-023-02260-y. J Cardiothorac Surg. 2023. PMID: 37038197 Free PMC article. Review.
-
Impact of patient-prosthesis mismatch on 30-day outcomes in young and middle-aged patients undergoing aortic valve replacement.J Cardiothorac Surg. 2012 May 15;7:46. doi: 10.1186/1749-8090-7-46. J Cardiothorac Surg. 2012. PMID: 22587627 Free PMC article.
-
Clinical predictors of prosthesis-patient mismatch after aortic valve replacement for aortic stenosis.Clinics (Sao Paulo). 2012;67(1):55-60. doi: 10.6061/clinics/2012(01)09. Clinics (Sao Paulo). 2012. PMID: 22249481 Free PMC article.
-
Patient Prosthesis Mismatch After SAVR and TAVR.Front Cardiovasc Med. 2022 Mar 30;9:761917. doi: 10.3389/fcvm.2022.761917. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 35433878 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical