An integrated macro/micro approach to evaluating pivot flow within the Medtronic ADVANTAGE bileaflet mechanical heart valve
- PMID: 12918854
An integrated macro/micro approach to evaluating pivot flow within the Medtronic ADVANTAGE bileaflet mechanical heart valve
Abstract
Background and aim of the study: An integrated macro/micro approach was used to evaluate flow within the pivots of the Medtronic ADVANTAGE bileaflet heart valve. Results were compared with those obtained with the St. Jude Medical bileaflet heart valve.
Methods: The integrated macro/micro approach consists of both a macroscopic hydrodynamic performance assessment and a three-part microscopic flow analysis. The hydrodynamic performance assesses the basic dynamic functions of the valves, while the microscopic flow analysis uses pivot flow visualization, computational fluid dynamics and laser Doppler velocimetry to determine pivot flow characteristics. Pivot flow visualization captures two-dimensional images of the pivot flow, defines the computational fluid dynamics boundary conditions, and validates the computational result. Three-dimensional unsteady computational fluid dynamics simulation reconstructs pivot flow structures. Laser Doppler velocimetry maps pivot velocity field and provides velocity validation for the computational simulation.
Results: The macroscopic hydrodynamic performance assessment showed the ADVANTAGE and St. Jude Medical valves to be comparable under identical flow conditions. The three techniques in the microscopic analysis mutually confirmed that the pivot design of the ADVANTAGE valve permits continuous-flow washing in the pivot recess, the pivots of both valves are completely wiped twice in a cardiac cycle, and no persistent pivot flow stases are observed.
Conclusion: The integrated macro/micro approach represents a powerful systematic method for determining detailed microscopic flow structures inside the pivots of bileaflet mechanical valves. The use of this technique during the design process of a bileaflet valve can eliminate the persistent flow stases that lead to thrombus formation.
Similar articles
-
Flow characterization of the ADVANTAGE and St. Jude Medical bileaflet mechanical heart valves.J Heart Valve Dis. 2004 Sep;13(5):814-22. J Heart Valve Dis. 2004. PMID: 15473485
-
The effect of gap width on viscous stresses within the leakage across a bileaflet valve pivot.J Heart Valve Dis. 2008 May;17(3):309-16. J Heart Valve Dis. 2008. PMID: 18592928
-
Computational fluid dynamics study of a protruded-hinge bileaflet mechanical heart valve.J Heart Valve Dis. 2001 Mar;10(2):254-262; discussion 263. J Heart Valve Dis. 2001. PMID: 11297213
-
Flow in prosthetic heart valves: state-of-the-art and future directions.Ann Biomed Eng. 2005 Dec;33(12):1689-94. doi: 10.1007/s10439-005-8759-z. Ann Biomed Eng. 2005. PMID: 16389514 Review.
-
Wear assessment in bileaflet heart valves.J Heart Valve Dis. 1996 Jun;5 Suppl 1:S133-43; discussion 144-8. J Heart Valve Dis. 1996. PMID: 8803766 Review.
Cited by
-
Fluid mechanics of artificial heart valves.Clin Exp Pharmacol Physiol. 2009 Feb;36(2):225-37. doi: 10.1111/j.1440-1681.2008.05099.x. Clin Exp Pharmacol Physiol. 2009. PMID: 19220329 Free PMC article. Review.
-
Clinical and echocardiographic assessment of the Medtronic Advantage aortic valve prosthesis: the Scandinavian multicentre, prospective study.Heart. 2007 Apr;93(4):500-5. doi: 10.1136/hrt.2005.086041. Epub 2006 Oct 25. Heart. 2007. PMID: 17065186 Free PMC article.
-
Simulation of the three-dimensional hinge flow fields of a bileaflet mechanical heart valve under aortic conditions.Ann Biomed Eng. 2010 Mar;38(3):841-53. doi: 10.1007/s10439-009-9857-0. Epub 2009 Dec 4. Ann Biomed Eng. 2010. PMID: 19960368 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Research Materials