Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator
- PMID: 12919860
- DOI: 10.1016/S0736-0266(02)00240-1
Polyethylene cross-linking by two different methods reduces acetabular liner wear in a hip joint wear simulator
Abstract
Advances in cross-linking have led to the development of wear resistant ultrahigh molecular weight polyethylene for total joint replacement. This study compared wear reduction by two different cross-linking methods as measured in a hip wear simulator. One highly cross-linked polyethylene was treated with 7.5 Mrad gamma irradiation with post-irradiation annealing and a sterilization dose of 2.5 Mrad (10 Gamma), while the other used 9.5 Mrad warm irradiation with 10 MeV electron-beam (9.5 EB). Liners of the same design, made from nominally cross-linked (gamma sterilized) polyethylene were also tested. Gravimetric wear analysis was performed every 500,000 cycles for 5,000,000 cycles. After correcting for weight gain due to water absorption, the nominally cross-linked liners demonstrated mean wear rates of 15.7 (+/-1.7) and 12.5 (+/-1.0) mg/million cycles. Both highly cross-linked polyethylene liners demonstrated significantly less wear than their respective controls (with mean wear rates of 1.5 (+/-1.2) and -1.4 (+/-1.5) mg/million cycles). The 9.5 EB liners gained weight presumably due to increased fluid absorption, in addition to that measured in loaded-soaked control implants. Any wear occurring was therefore assumed to have been more than offset by weight gain. Highly cross-linked polyethylene was significantly more wear resistant than non- or nominally cross-linked polyethylene. The differences in wear rates between the two highly cross-linked polyethylene designs (9.5 EB or 10 Gamma) are probably too small to be clinically significant.
Similar articles
-
Wear of conventional and cross-linked ultra-high-molecular-weight polyethylene acetabular cups against polished and roughened CoCr femoral heads in a biaxial hip simulator.J Biomed Mater Res. 2002;63(6):848-53. doi: 10.1002/jbm.10471. J Biomed Mater Res. 2002. PMID: 12418033
-
Impingement of acetabular cups in a hip simulator: comparison of highly cross-linked and conventional polyethylene.J Arthroplasty. 2005 Oct;20(7 Suppl 3):77-86. doi: 10.1016/j.arth.2005.04.026. J Arthroplasty. 2005. PMID: 16214007
-
The impact of storage time on the wear rates of ultrahigh-molecular-weight polyethylene acetabular liners in hip simulators.J Arthroplasty. 2009 Jun;24(4):543-8. doi: 10.1016/j.arth.2008.01.308. Epub 2008 Apr 11. J Arthroplasty. 2009. PMID: 18534409
-
Conventional ultra-high molecular weight polyethylene: a gold standard of sorts.Instr Course Lect. 2005;54:183-7. Instr Course Lect. 2005. PMID: 15948445 Review.
-
Highly cross-linked polyethylene bearing surfaces in total hip arthroplasty.J Surg Orthop Adv. 2008 Spring;17(1):27-33. J Surg Orthop Adv. 2008. PMID: 18284901 Review.
Cited by
-
The John Charnley Award: an accurate and sensitive method to separate, display, and characterize wear debris: part 1: polyethylene particles.Clin Orthop Relat Res. 2012 Feb;470(2):329-38. doi: 10.1007/s11999-011-2057-x. Clin Orthop Relat Res. 2012. PMID: 21997780 Free PMC article.
-
Design features of the rotating head total hip arthroplasty by Weber-Huggler- a forgotten technical solution to reduce wear.Arch Orthop Trauma Surg. 2025 Jan 23;145(1):140. doi: 10.1007/s00402-025-05768-1. Arch Orthop Trauma Surg. 2025. PMID: 39849197 Review.
-
Head-neck taper corrosion in hip arthroplasty.Biomed Res Int. 2015;2015:758123. doi: 10.1155/2015/758123. Epub 2015 Apr 14. Biomed Res Int. 2015. PMID: 25954757 Free PMC article. Review.
-
CMM-Based Volumetric Assessment Methodology for Polyethylene Tibial Knee Inserts in Total Knee Replacement.Appl Bionics Biomech. 2018 Apr 10;2018:9846293. doi: 10.1155/2018/9846293. eCollection 2018. Appl Bionics Biomech. 2018. PMID: 29849764 Free PMC article.
-
Equivalent wear performance of dual mobility bearing compared with standard bearing in total hip arthroplasty: in vitro study.Int Orthop. 2017 Mar;41(3):521-527. doi: 10.1007/s00264-016-3346-5. Epub 2016 Nov 22. Int Orthop. 2017. PMID: 27878334
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources