Clinical Trial
. 2003 Sep;89(9):1014-8.
doi: 10.1136/heart.89.9.1014.

Portable spectral Doppler echocardiographic device: overcoming limitations

Affiliations
Clinical Trial

Portable spectral Doppler echocardiographic device: overcoming limitations

J Quiles et al. Heart. 2003 Sep.

Abstract

Background: There is evidence that new portable echocardiographic devices are useful in evaluating heart anatomy and function, but a lack of Doppler modes has up to now been an important limitation in obtaining haemodynamic data.

Objectives: To report the Doppler capabilities of a new hand held echocardiographic device.

Design: Blinded comparison of two types of echocardiography machine.

Setting: Tertiary care centre.

Patients: 98 consecutive patients were randomly imaged with the hand held device, with a standard platform as reference.

Outcome measures: Pulsed wave transmitral Doppler inflow tract velocities, deceleration time, and continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation peak velocities were recorded.

Results: There was excellent agreement between the hand held device and standard echocardiography for the evaluation of diastolic E and A waves, E/A ratio, and deceleration time with pulsed wave Doppler (intraclass correlation coefficients of 0.97, 0.93, 0.90, and 0.78, respectively). In addition, good agreement was found between continuous wave Doppler measurements of aortic ejection and tricuspid regurgitation velocities (intraclass correlation coefficients of 0.96 and 0.80). However, there was a significant difference between patients with tricuspid regurgitation measured with the hand held device (25.5%) and by standard echocardiography (65.3%), resulting in misdiagnosis of eight patients with pronounced pulmonary hypertension.

Conclusions: New hand held devices with Doppler capabilities overcome previous limitations in evaluating haemodynamic variables. With colour Doppler they are now suitable for the complete evaluation of valvar disease and diastolic function. However, important limitations remain in the evaluation of pulmonary pressures.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Bland-Altman plot, showing the differences between pulsed wave Doppler measurements by hand held echocardiography and standard echocardiography. The mean of the differences and 1.96 standard deviations (SD) of the mean difference are shown.
Figure 2
Figure 2
Bland-Altman plot showing the magnitude of the difference between continuous Doppler measurement of aortic maximum velocity and tricuspid regurgitation velocity measured by hand held echocardiography and standard echocardiography. 1.96 (SD), 1.96 standard deviations of the mean difference.

Comment in

References

    1. Spencer KT, Anderson AS, Bhargava A, et al. Physician-performed point-of-care echocardiography using a laptop platform compared with physical examination in the cardiovascular patient. J Am Coll Cardiol 2001;37:2013–18. - PubMed
    1. Rugolotto M, Hu BS, Liang DH, et al. Rapid assessment of cardiac anatomy and function with a new hand-carried ultrasound device (OptiGo): a comparison with standard echocardiography. Eur J Echocardiogr 2001;2:262–9. - PubMed
    1. Vourvouri E, Poldermans D, Schinkel A, et al. Left ventricular hypertrophy screening using a hand-held ultrasound device. Eur Heart J 2002;23:1516–21. - PubMed
    1. Vourvouri EC, Poldermans D, Schinkel AF, et al. Abdominal aortic aneurysm screening using a hand-held ultrasound device. “A pilot study”. Eur J Vasc Endovasc Surg 2001;22:352–4. - PubMed
    1. Vourvouri EC, Poldermans D, De Sutter J, et al. Experience with an ultrasound stethoscope. J Am Soc Echocardiogr 2002;15:80–5. - PubMed

Publication types

MeSH terms