Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube
- PMID: 12923620
- DOI: 10.1007/s00134-003-1933-6
Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube
Abstract
Objective: This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-microm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff.
Design: In vitro setup.
Measurements and results: The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30 cmH2O.
Conclusions: Within the acceptable upper limit for tracheal cuff pressure (25-30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.
Similar articles
-
Air leakage around endotracheal tube cuffs.Eur J Anaesthesiol. 2004 Jun;21(6):448-53. doi: 10.1017/s0265021504006064. Eur J Anaesthesiol. 2004. PMID: 15248624 Clinical Trial.
-
Nitrous oxide diffusion into tracheal tube cuffs: comparison of five different tracheal tube cuffs.Acta Anaesthesiol Scand. 2004 Oct;48(9):1180-4. doi: 10.1111/j.1399-6576.2004.00483.x. Acta Anaesthesiol Scand. 2004. PMID: 15352966
-
Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study.Br J Anaesth. 2010 Oct;105(4):538-43. doi: 10.1093/bja/aeq202. Epub 2010 Aug 3. Br J Anaesth. 2010. PMID: 20682571
-
Tracheal cuff pressure monitoring in the ICU: a literature review and survey of current practice in Queensland.Anaesth Intensive Care. 2014 Nov;42(6):761-70. doi: 10.1177/0310057X1404200612. Anaesth Intensive Care. 2014. PMID: 25342409 Review.
-
[Endotracheal tubes in pediatric patients. Published formulas to estimate the optimal size].Anaesthesist. 2011 Apr;60(4):334-42. doi: 10.1007/s00101-010-1756-0. Anaesthesist. 2011. PMID: 20593158 Review. German.
Cited by
-
Endotracheal tube cuff pressure monitoring during neurosurgery - Manual vs. automatic method.J Anaesthesiol Clin Pharmacol. 2011 Jul;27(3):358-61. doi: 10.4103/0970-9185.83682. J Anaesthesiol Clin Pharmacol. 2011. PMID: 21897508 Free PMC article.
-
Pediatric cuffed endotracheal tubes: an evolution of care.Ochsner J. 2011 Spring;11(1):52-6. Ochsner J. 2011. PMID: 21603336 Free PMC article.
-
Randomized Pilot Trial of Two Modified Endotracheal Tubes To Prevent Ventilator-associated Pneumonia.Ann Am Thorac Soc. 2016 Jan;13(1):72-80. doi: 10.1513/AnnalsATS.201506-346OC. Ann Am Thorac Soc. 2016. PMID: 26523433 Free PMC article. Clinical Trial.
-
CO2 driven endotracheal tube cuff control in critically ill patients: A randomized controlled study.PLoS One. 2017 May 11;12(5):e0175476. doi: 10.1371/journal.pone.0175476. eCollection 2017. PLoS One. 2017. PMID: 28493877 Free PMC article. Clinical Trial.
-
A polyurethane cuffed endotracheal tube is associated with decreased rates of ventilator-associated pneumonia.J Crit Care. 2011 Jun;26(3):280-6. doi: 10.1016/j.jcrc.2010.05.035. Epub 2010 Jul 23. J Crit Care. 2011. PMID: 20655698 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous