. 2003 Oct;29(10):1849-53.
doi: 10.1007/s00134-003-1933-6. Epub 2003 Aug 16.

Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube

Affiliations

Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube

Alexander Dullenkopf et al. Intensive Care Med. 2003 Oct.

Abstract

Objective: This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-microm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff.

Design: In vitro setup.

Measurements and results: The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10-30 cmH2O.

Conclusions: Within the acceptable upper limit for tracheal cuff pressure (25-30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Anaesthesia. 2002 Feb;57(2):173-5 - PubMed
    1. Br J Anaesth. 1992 May;68(5):499-502 - PubMed
    1. Br J Anaesth. 2002 Mar;88(3):456; author reply 456-7 - PubMed
    1. Br J Anaesth. 1997 May;78(5):557-62 - PubMed
    1. Br Med J (Clin Res Ed). 1984 Mar 31;288(6422):965-8 - PubMed

Publication types

MeSH terms

LinkOut - more resources