Comparative Study
. 2003 Aug;7(4):R67.
doi: 10.1186/cc2339. Epub 2003 Jun 11.

Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Affiliations
Comparative Study

Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?

Gavin D Perkins et al. Crit Care. 2003 Aug.

Abstract

Introduction: This study investigates the relation between changes in pulse oximeter oxygen saturation (SpO2) and changes in arterial oxygen saturation (SaO2) in the critically ill, and the effects of acidosis and anaemia on precision of using pulse oximetry to predict SaO2.

Patients and methods: Forty-one consecutive patients were recruited from a nine-bed general intensive care unit into a 2-month study. Patients with significant jaundice (bilirubin >40 micromol/l) or inadequate pulse oximetry tracing were excluded.

Results: A total of 1085 paired readings demonstrated only moderate correlation (r= 0.606; P < 0.01) between changes in SpO2 and those in SaO2, and the pulse oximeter tended to overestimate actual changes in SaO2. Anaemia increased the degree of positive bias whereas acidosis reduced it. However, the magnitude of these changes was small.

Conclusion: Changes in SpO2 do not reliably predict equivalent changes in SaO2 in the critically ill. Neither anaemia nor acidosis alters the relation between SpO2 and SaO2 to any clinically important extent.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Linear relations between changes in pulse oximeter oxygen saturation (SpO2) and arterial oxygen saturation (SaO2).
Figure 2
Figure 2
Bland and Altman plot for bias and limits of agreement for total data. SaO2, arterial oxygen saturation; SpO2, pulse oximeter oxygen saturation.

References

    1. Jubran A. Advances in respiratory monitoring during mechanical ventilation. Chest. 1999;116:1416–1425. doi: 10.1378/chest.116.5.1416. - DOI - PubMed
    1. Paky F, Koeck CM. Pulse oximetry in ventilated preterm newborns: reliability of detection of hyperoxaemia and hypoxaemia, and feasibility of alarm settings. Acta Paediatr. 1995;84:613–616. - PubMed
    1. Inman KJ, Sibbald WJ, Rutledge FS, Speechley M, Martin CM, Clark BJ. Does implementing pulse oximetry in a critical care unit result in substantial arterial blood gas savings? Chest. 1993;104:542–546. - PubMed
    1. Jubran A, Tobin MJ. Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990;97:1420–1425. - PubMed
    1. Slutsky AS. Consensus conference on mechanical ventilation: January 28–30, 1993 at Northbrook, Illinois, USA. Part I. European Society of Intensive Care Medicine, the ACCP and the SCCM. Intensive Care Med. 1994;20:64–79. - PubMed

Publication types