Comparison of two intrauterine pressure catheters during labor
- PMID: 12953323
Comparison of two intrauterine pressure catheters during labor
Abstract
Objective: To compare the Koala air-charged coupled catheter (Clinical Innovations, Inc., Murray, Utah) for monitoring intrauterine pressure (IUP) during active labor with an electronic pressure transducer-tip catheter, Intran Plus (Utah Medical Products, Midvale, Utah).
Study design: Simultaneous recordings with Koala and Intran catheters bound together at their tips were made in 18 term patients using comparably calibrated fetal monitors.
Results: Coefficients of correlation between the two IUP catheter systems for mean uterine baseline, peak pressures, contraction frequency and duration were .78, .90, 1.00 and .63, respectively. Both systems functioned equally well in displaying abnormal uterine activity. Some difficulty was encountered in placing the combined catheters in the amniotic fluid space: 4 of 18 (23%) were placed outside the chorionic membrane, in the extraovular space.
Conclusion: Comparisons of the recordings from the Koala and Intran Plus catheters showed similar mean baseline uterine tone, peak pressures, contraction frequency and duration.
Similar articles
-
Evaluating a new intrauterine pressure catheter.J Reprod Med. 1997 Aug;42(8):506-13. J Reprod Med. 1997. PMID: 9284013
-
Reliability of intrauterine pressure measurements.Obstet Gynecol. 1991 Nov;78(5 Pt 1):800-2. Obstet Gynecol. 1991. PMID: 1923200
-
Monitoring intrauterine pressure during active labor. A prospective comparison of two methods.J Reprod Med. 1989 Oct;34(10):811-4. J Reprod Med. 1989. PMID: 2795563
-
Uterine activity monitoring during labor.J Perinat Med. 2007;35(6):468-77. doi: 10.1515/JPM.2007.116. J Perinat Med. 2007. PMID: 18052832 Review.
-
Monitoring uterine activity.Br J Hosp Med. 1993 May 5-18;49(9):649-53. Br J Hosp Med. 1993. PMID: 8508259 Review.
Cited by
-
Estimation of Physiologic Pressures: Invasive and Non-Invasive Techniques, AI Models, and Future Perspectives.Sensors (Basel). 2023 Jun 20;23(12):5744. doi: 10.3390/s23125744. Sensors (Basel). 2023. PMID: 37420919 Free PMC article. Review.
-
Comparison of air-coupled balloon esophageal and anorectal manometry catheters with solid-state esophageal manometry and water-perfused anorectal manometry catheters.Dig Dis Sci. 2004 Oct;49(10):1657-63. doi: 10.1023/b:ddas.0000043382.59539.d3. Dig Dis Sci. 2004. PMID: 15573923 Clinical Trial.
-
Labour diagnostics using the intrauterine pressure through sound waves emitted by a smartphone.Med Biol Eng Comput. 2025 Jun;63(6):1849-1866. doi: 10.1007/s11517-025-03305-1. Epub 2025 Jan 31. Med Biol Eng Comput. 2025. PMID: 39888469 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical