[Film-screen mammography versus digital storage plate mammography: hard copy and monitor display of microcalcifications and focal findings--a retrospective clinical and histologic analysis]

[Article in German]
Affiliations
Comparative Study

[Film-screen mammography versus digital storage plate mammography: hard copy and monitor display of microcalcifications and focal findings--a retrospective clinical and histologic analysis]

[Article in German]
R Schulz-Wendtland et al. Rofo. 2003 Sep.

Abstract

Purpose: A retrospective clinical-histological study to determine the diagnostic accuracy of mammography using conventional screen-film cassettes (hard copy), high-resolution digital phosphor storage plates (hard copy) and monitor display (soft copy) for microcalcifications and focal lesions (BI-RADS (TM) category 4 or 5).

Materials and methods: From April to November 2001, 76 patients underwent conventional film-screen mammography and, after diagnosis and preoperative wire localization, digital mammography with the same exposure parameters. Five investigators retrospectively determined the diagnosis after the operation from randomly distributed mediolateral views (hard-copy reading) and from the monitor display (soft-copy reading). These results were correlated with the final histology.

Results: The accuracy of conventional screen-film mammography, digital mammography and monitor-displayed mammography was 67%, 65% and 68% for all findings, (n = 76), 59%, 59% and 68% for microcalcifications (n = 44) and 75%, 72% and 63% for focal lesions (n = 32). The overall results showed no difference.

Conclusion: Our findings indicate equivalence of conventional screen-film mammography, high-resolution digital phosphor storage plate mammography and monitor-displayed mammography.