Pulse oximetry during labour--does it give rise to hope? Value of saturation monitoring in comparison to fetal blood gas status
- PMID: 12965102
- DOI: 10.1016/s0301-2115(03)00184-2
Pulse oximetry during labour--does it give rise to hope? Value of saturation monitoring in comparison to fetal blood gas status
Abstract
Objective: Purpose of this presentation is to show the diagnostic power of fetal pulse oximetry in comparison to the other blood gas parameters from fetal blood samplings (FBS). The distribution of saturation readings in acidotic fetuses and normally oxygenated fetuses should be established.
Study design: A fetal pulse oximetry system (N400, FS14) was evaluated in a strictly observational study design based on blinded saturation display and on continuous data storing. The investigation was performed on 170 fetuses with non-reassuring fetal heart rate (FHR)-tracings. Since pulse oximetry readings were not available for decision finding, the clinical management was based on electronic fetal monitoring and fetal blood samplings. The oxygen saturation from FBS or umbilical cord blood was measured by blood gas analysers with an integrated hemoximeter (Bayer 865; ABL 625, Radiometer) and biosensors measuring lactate as metabolic component. Out of the 170 cases 17 cases were defined as group of acidemia (pH(umb.art.) < 7.16 + BD<-9.4). The distribution of saturation readings and the duration of desaturation periods (in minutes and percentage of total monitoring time SPO2 below 30%) were determined. ROC curve analysis from FBS preceding delivery compared the diagnostic power of other blood gas parameters with oxygen saturation. The Wilcoxon test for uneven pairs was used.
Results: The distribution of oxygen saturation in the normal group of fetuses differs significantly from the acidemic group. The correlation coefficient between both methods to determine oxygen saturation was r=0.66. A specific evaluation of the distribution of SPO2 shows an overestimation of pulse oximetry in the low range and an underestimation in the high range of saturation. ROC-curve analysis showed a good diagnostic power of lactate in comparison to the oxygen saturation measured by pulse oximetry or by hemoximetry.
Conclusions: The advantage of continuous fetal pulse oximetry surveilling the fetus under suspicion of hypoxia appears limited by the poor diagnostic power of the respiratory parameter saturation itself and by the impairment of the precision of the technology.
Similar articles
-
Accuracy of fetal pulse oximetry.Acta Obstet Gynecol Scand. 2002 May;81(5):417-23. doi: 10.1034/j.1600-0412.2002.810508.x. Acta Obstet Gynecol Scand. 2002. PMID: 12027815
-
[Outcome measures in perinatal medicine--pO2 and SO2. With remarks on pulse oximetry].Z Geburtshilfe Neonatol. 2005 Oct;209(5):173-85. doi: 10.1055/s-2005-873203. Z Geburtshilfe Neonatol. 2005. PMID: 16317628 Clinical Trial. German.
-
The prediction of fetal acidosis by means of intrapartum fetal pulse oximetry.Am J Obstet Gynecol. 1999 Jan;180(1 Pt 1):73-81. doi: 10.1016/s0002-9378(99)70153-5. Am J Obstet Gynecol. 1999. PMID: 9914582
-
Continuous intrapartum pH, pO2, pCO2, and SpO2 monitoring.Obstet Gynecol Clin North Am. 1999 Dec;26(4):671-93. doi: 10.1016/s0889-8545(05)70106-6. Obstet Gynecol Clin North Am. 1999. PMID: 10587962 Review.
-
[Fetal scalp blood sampling for pH and lactate measurement during labour].J Gynecol Obstet Biol Reprod (Paris). 2008 Feb;37 Suppl 1:S65-71. doi: 10.1016/j.jgyn.2007.11.012. Epub 2008 Jan 7. J Gynecol Obstet Biol Reprod (Paris). 2008. PMID: 18179877 Review. French.
Cited by
-
Association between intrapartum fetal pulse oximetry and adverse perinatal and long-term outcomes: a systematic review and meta-analysis protocol.HRB Open Res. 2024 Mar 28;6:63. doi: 10.12688/hrbopenres.13802.2. eCollection 2023. HRB Open Res. 2024. PMID: 38628596 Free PMC article.
-
Haemodynamic Instability and Brain Injury in Neonates Exposed to Hypoxia⁻Ischaemia.Brain Sci. 2019 Feb 27;9(3):49. doi: 10.3390/brainsci9030049. Brain Sci. 2019. PMID: 30818842 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous