Blinded multi-laboratory evaluation of an in-house dot-blot ELISA kit for diagnosis of human parastrongyliasis
- PMID: 12971507
Blinded multi-laboratory evaluation of an in-house dot-blot ELISA kit for diagnosis of human parastrongyliasis
Abstract
An in-house membrane dot-blot enzyme-linked immunosorbent assay kit, for the detection of specific antibody to a 31 kDa protein of Parastrongylus cantonensis was evaluated in a blinded multi-laboratory study. The kit was provided to technologists, who were trained in its use, from nine regional laboratories of Thailand. With an identical set of 20 coded serum samples and reference positive and negative controls, the kit was found to have a diagnostic sensitivity of 100% and a diagnostic specificity of 100% in all the laboratories. There was no obvious variation in quality among five lots of the antigen-coated nitrocellulose strips evaluated. The shelf life of the kit was > or = 6 months when the test components were stored at 4 degrees C and 18 months at -20 degrees C. The enzyme-linked immunoassay dot technique is easy to perform and does not require sophisticated electrical equipment; the result is available within 3 hours. If appropriate technical training is included, the application of this dot-blot ELISA kit in clinical laboratories throughout Thailand should be possible.
Similar articles
-
Detection of circulating antigens of Parastrongylus cantonensis in human sera by dot-blot ELISA and sandwich ELISA using monoclonal antibody.Southeast Asian J Trop Med Public Health. 1997 Sep;28(3):624-8. Southeast Asian J Trop Med Public Health. 1997. PMID: 9561620
-
Purification of a specific immunodiagnostic Parastrongylus cantonensis antigen by electroelution from SDS-polyacrylamide gels.Southeast Asian J Trop Med Public Health. 2001 Jun;32(2):308-13. Southeast Asian J Trop Med Public Health. 2001. PMID: 11556581
-
Multi-immunodot for rapid differential diagnosis of eosinophilic meningitis due to parasitic infections.J Helminthol. 2006 Sep;80(3):249-54. J Helminthol. 2006. PMID: 16923267
-
Immunological diagnosis of human angiostrongyliasis due to Angiostrongylus cantonensis (Nematoda: Angiostrongylidae).Int J Infect Dis. 2009 Jul;13(4):425-31. doi: 10.1016/j.ijid.2008.09.021. Epub 2008 Dec 30. Int J Infect Dis. 2009. PMID: 19117782 Review.
-
Third report of ocular parastrongyliasis (angiostrongyliasis) from Sri Lanka.Parassitologia. 2001 Sep;43(3):95-7. Parassitologia. 2001. PMID: 11921545 Review.
Cited by
-
Update on eosinophilic meningoencephalitis and its clinical relevance.Clin Microbiol Rev. 2009 Apr;22(2):322-48, Table of Contents. doi: 10.1128/CMR.00044-08. Clin Microbiol Rev. 2009. PMID: 19366917 Free PMC article. Review.
-
Immunodiagnostic Detection of Angiostrongylus cantonensis Exposure on Hawaii Island Using Isogeographic 31-kDa Antigen.Am J Trop Med Hyg. 2023 Jun 12;109(2):335-342. doi: 10.4269/ajtmh.22-0643. Print 2023 Aug 2. Am J Trop Med Hyg. 2023. PMID: 37308099 Free PMC article.
-
Anti-N-methyl-D-aspartate receptor encephalitis associated with intracranial Angiostrongylus cantonensis infection: a case report.Neurol Sci. 2017 Apr;38(4):703-706. doi: 10.1007/s10072-016-2718-3. Epub 2016 Oct 24. Neurol Sci. 2017. PMID: 27778112
-
Estimating Human Exposure to Rat Lungworm (Angiostrongylus cantonensis) on Hawai'i Island: A Pilot Study.Am J Trop Med Hyg. 2020 Jan;102(1):69-77. doi: 10.4269/ajtmh.18-0242. Am J Trop Med Hyg. 2020. PMID: 31769399 Free PMC article.
-
The current status of laboratory diagnosis of Angiostrongylus cantonensis infections in humans using serologic and molecular methods.Hawaii J Med Public Health. 2013 Jun;72(6 Suppl 2):55-7. Hawaii J Med Public Health. 2013. PMID: 23901386 Free PMC article.