. 1992 Jan;61(1):78-82.
doi: 10.1016/S0006-3495(92)81817-9.

Novel voltage clamp to record small, fast currents from ion channels expressed in Xenopus oocytes

Affiliations

Novel voltage clamp to record small, fast currents from ion channels expressed in Xenopus oocytes

M Taglialatela et al. Biophys J. 1992 Jan.

Abstract

The present report describes a novel technique for voltage-clamping amphibian oocytes in which part of the membrane is isolated by a vaseline gap and the cytoplasmic fluid is exchanged by cutting or permeabilizing the remaining membrane. The main features of this open-oocyte, vaseline-gap voltage clamp are: (a) low current noise (1 nA at 3 kHz), (b) control of the ionic composition of both the internal and external media, (c) fast time resolution (20-100 microseconds time constant of decay of the capacity transient) and (d) stable recordings for several hours. These features allow reliable measurements of tail or gating currents and the new method is especially suitable when either of these currents must be measured to test the effects of mutations introduced into the cDNAs of cloned ion channels.

PubMed Disclaimer

References

    1. Nature. 1988 Jan 14;331(6152):143-5 - PubMed
    1. FEBS Lett. 1988 Feb 8;228(1):195-200 - PubMed
    1. J Gen Physiol. 1989 Nov;94(5):953-69 - PubMed
    1. J Physiol. 1984 Dec;357:173-83 - PubMed
    1. J Physiol. 1983 Sep;342:309-25 - PubMed

Publication types