Signal delivery/real ear measurement system for hearing aid selection and fitting
- PMID: 1342584
Signal delivery/real ear measurement system for hearing aid selection and fitting
Abstract
This report describes a signal delivery/real ear measurement system for application in the hearing aid selection and fitting process. This signal delivery/real ear measurement system provides a means for quantifying a listener's auditory characteristics (e.g., thresholds, loudness discomfort levels) in a manner that is compatible with electroacoustic measures of hearing aid performance. The signal delivery/real ear measurement system consists of a button-type hearing aid receiver coupled to the listener's personal earmold, and a probe tube microphone system to measure the signal level within the occluded ear canal. This signal delivery/real ear measurement system was used to measure detection thresholds and loudness discomfort levels in severe/profoundly hearing-impaired school-age children, with results indicating good test-retest reliability in behavioral responses. The findings presented in this report relate to the intersession electroacoustic variability associated with this instrumentation. In addition, the potential application of this or a similar system for measuring real ear to 2-cc coupler difference values is illustrated and discussed.
Comment in
-
Within-subject versus between-subject variability comparison.Ear Hear. 1993 Aug;14(4):293-4. doi: 10.1097/00003446-199308000-00009. Ear Hear. 1993. PMID: 8405734 No abstract available.
Similar articles
-
Comparison of real-ear to coupler difference values in the right and left ear of hearing aid users.Ear Hear. 2010 Feb;31(1):146-50. doi: 10.1097/AUD.0b013e3181b8399b. Ear Hear. 2010. PMID: 19745743
-
Real-ear measurement verification for open, non-occluding hearing instruments.Int J Audiol. 2007 Jan;46(1):11-6. doi: 10.1080/14992020601083313. Int J Audiol. 2007. PMID: 17365050
-
Behavioral audiometry: protocols for measuring hearing thresholds in babies aged 4-18 months.Int J Pediatr Otorhinolaryngol. 2004 Oct;68(10):1233-43. doi: 10.1016/j.ijporl.2004.04.030. Int J Pediatr Otorhinolaryngol. 2004. PMID: 15364493
-
Bilateral digital hearing aids for binaural hearing.Ear Nose Throat J. 1994 Mar;73(3):176-9. Ear Nose Throat J. 1994. PMID: 8205979 Review.
-
[In-situ measurement as necessary component of hearing aid fitting].Laryngorhinootologie. 1991 Nov;70(11):616-9. doi: 10.1055/s-2007-998109. Laryngorhinootologie. 1991. PMID: 1755901 Review. German.
Cited by
-
Clinical protocols for hearing instrument fitting in the Desired Sensation Level method.Trends Amplif. 2005;9(4):199-226. doi: 10.1177/108471380500900404. Trends Amplif. 2005. PMID: 16424946 Free PMC article. Review.
-
An approach for ensuring accuracy in pediatric hearing instrument fitting.Trends Amplif. 2003;7(1):29-40. doi: 10.1177/108471380300700104. Trends Amplif. 2003. PMID: 15004646 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical