Effect of fetal haemoglobin on the accuracy of pulse oximetry in preterm infants
- PMID: 1372809
- DOI: 10.1111/j.1440-1754.1992.tb02615.x
Effect of fetal haemoglobin on the accuracy of pulse oximetry in preterm infants
Abstract
Pulse oximeters are programmed with a calibration curve derived from studies done in adults. Whether fetal haemoglobin levels affect their reliability is unclear. This study reports the accuracy of pulse oximetry in 22 preterm infants (mean 31 weeks, range 25-36 weeks gestation) between 1 h and 73 days of age. Oxygen saturation obtained from a Nellcor N-200 pulse oximeter (SpO2) was compared with simultaneous arterial values (functional SaO2) measured by a Radiometer OSM3 Hemoximeter over a SpO2 range of 83-99%. Fetal haemoglobin (HbF), carboxyhaemoglobin (HbCO) and methaemoglobin (HbMet) measured by the hemoximeter ranged between 0-100%, 0-3.5% and 0-0.8% respectively. Linear regression analysis revealed a close correlation between SpO2 and functional SaO2 (SpO2 = 0.75 SaO2 + 24.43, r = 0.88, P less than 0.001) over a wide range of values for PCV, heart rate, blood pressure, PaO2, PaCO2 and pH. The mean SpO2-SaO2 difference of 1.3, (s.d. 2.5%, P less than 0.001) was unaffected by HgF, HbCO or HbMet but was increased in infants receiving inotropic support. We conclude that the Nellcor N-200 pulse oximeter gives reliable oxygen saturation measurements unaffected by the HbF level in preterm infants.
Similar articles
-
Accuracy of Pulse Oximetry in the Presence of Fetal Hemoglobin-A Systematic Review.Children (Basel). 2021 Apr 30;8(5):361. doi: 10.3390/children8050361. Children (Basel). 2021. PMID: 33946236 Free PMC article. Review.
-
Saturation by pulse oximetry: comparison of the results obtained by instruments of different brands.J Pediatr. 1993 Apr;122(4):620-6. doi: 10.1016/s0022-3476(05)83549-8. J Pediatr. 1993. PMID: 7681875
-
Pulse oximetry in newborn infants with birth weights of 620 to 4285 grams receiving dopamine and dobutamine.J Perinatol. 1996 Jan-Feb;16(1):31-4. J Perinatol. 1996. PMID: 8869537
-
Pulse oximetry in very low birth weight infants with acute and chronic lung disease.Pediatrics. 1987 Apr;79(4):612-7. Pediatrics. 1987. PMID: 2434913
-
From oxygen content to pulse oximetry: completing the picture in the newborn.Acta Anaesthesiol Scand Suppl. 1995;107:95-100. doi: 10.1111/j.1399-6576.1995.tb04341.x. Acta Anaesthesiol Scand Suppl. 1995. PMID: 8599308 Review.
Cited by
-
Hemoglobin oxygen saturation targets in the neonatal intensive care unit: Is there a light at the end of the tunnel? 1.Can J Physiol Pharmacol. 2019 Mar;97(3):174-182. doi: 10.1139/cjpp-2018-0376. Epub 2018 Oct 26. Can J Physiol Pharmacol. 2019. PMID: 30365906 Free PMC article. Review.
-
Factors influencing fetal pulse oximetry performance.J Clin Monit Comput. 2004 Feb;18(1):13-24. doi: 10.1023/b:jocm.0000025278.82852.b3. J Clin Monit Comput. 2004. PMID: 15139579 Review.
-
Accuracy of Pulse Oximetry in the Presence of Fetal Hemoglobin-A Systematic Review.Children (Basel). 2021 Apr 30;8(5):361. doi: 10.3390/children8050361. Children (Basel). 2021. PMID: 33946236 Free PMC article. Review.
-
The measurement of accurate fetal hemoglobin and related oxygen saturation by the hemoximeter.Clin Chim Acta. 2006 Dec;374(1-2):75-80. doi: 10.1016/j.cca.2006.05.042. Epub 2006 Jun 6. Clin Chim Acta. 2006. PMID: 16844104 Free PMC article.
-
Oxygen targeting in preterm infants: a physiological interpretation.J Perinatol. 2015 Jan;35(1):8-15. doi: 10.1038/jp.2014.199. Epub 2014 Oct 30. J Perinatol. 2015. PMID: 25357098 Free PMC article. Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical