Alterations of phasic coronary artery flow velocity in humans during percutaneous coronary angioplasty
- PMID: 1386088
- DOI: 10.1016/0735-1097(92)90091-z
Alterations of phasic coronary artery flow velocity in humans during percutaneous coronary angioplasty
Abstract
Background and objectives: Studies using Doppler catheters to assess blood flow velocity and vasodilator reserve in proximal coronary arteries have failed to demonstrate significant improvement immediately after coronary angioplasty. Measurement of blood flow velocity, flow reserve and phasic diastolic/systolic velocity ratio performed distal to a coronary stenosis may provide important information concerning the physiologic significance of coronary artery stenosis. This study was designed to measure these blood flow velocity variables both proximal and distal to a significant coronary artery stenosis in patients undergoing coronary angioplasty.
Methods: A low profile (0.018-in.) (0.046-cm) Doppler angioplasty guide wire capable of providing spectral flow velocity data was used to measure blood flow velocity, flow reserve and diastolic/systolic velocity ratio both proximal and distal to left anterior descending or left circumflex coronary artery stenosis. These measurements were made in 38 patients undergoing coronary angioplasty and in 12 patients without significant coronary artery disease.
Results: Significant improvement in mean time average peak velocity was noted in distal coronary arteries after angioplasty (before 19 +/- 12 cm/s; after 35 +/- 16 cm/s; p less than 0.01). Increases in proximal average peak velocity after angioplasty were less remarkable (before 34 +/- 18 cm/s; after 41 +/- 14 cm/s; p = 0.04). Mean flow reserve remained unchanged after angioplasty both proximal (1.5 +/- 0.5 vs. 1.6 +/- 1; p greater than 0.10) and distal (1.6 +/- 1 vs. 1.5 +/- 0.8; p greater than 0.10) to a coronary stenosis. Before angioplasty, mean diastolic/systolic velocity ratio measured distal to a significant stenosis was decreased compared with that in normal vessels (1.3 +/- 0.5 vs. 1.8 +/- 0.5; p less than 0.01). After angioplasty, distal abnormal phasic velocity patterns generally returned to normal, with a significant increase in mean diastolic/systolic velocity ratio (1.3 +/- 0.5 vs. 1.9 +/- 0.6; p less than 0.01). Phasic velocity patterns and mean diastolic/systolic velocity ratio measured proximal to a coronary stenosis were not statistically different from values in normal vessels (1.8 +/- 0.8 vs. 1.8 +/- 0.5; p greater than 0.10) and did not change significantly after angioplasty (1.8 +/- 0.8 vs. 2.13 +/- 0.9; p greater than 0.10).
Conclusions: Flow velocity measurements may be performed distal to a coronary stenosis with the Doppler guide wire. Phasic velocity measurements made proximal to a coronary stenosis differed from those in the distal coronary artery. Both proximal and distal flow reserve measurements made immediately after angioplasty were of limited utility. Changes in distal flow velocity patterns and diastolic/systolic velocity ratio appeared to be more relevant than the hyperemic response in assessing the immediate physiologic outcome of coronary angioplasty.
Similar articles
-
Analysis of coronary blood flow velocity dynamics in angiographically normal and stenosed arteries before and after endolumen enlargement by angioplasty.J Am Coll Cardiol. 1993 Feb;21(2):308-16. doi: 10.1016/0735-1097(93)90668-q. J Am Coll Cardiol. 1993. PMID: 8425991
-
Blood flow velocity in the right coronary artery: assessment before and after angioplasty.J Am Coll Cardiol. 1994 Oct;24(4):1012-7. doi: 10.1016/0735-1097(94)90863-x. J Am Coll Cardiol. 1994. PMID: 7930191
-
Applications of coronary flow velocity during angioplasty and other coronary interventional procedures.Am J Cardiol. 1993 May 20;71(14):17D-25D. doi: 10.1016/0002-9149(93)90130-5. Am J Cardiol. 1993. PMID: 8488771 Review.
-
Use of a low profile Doppler angioplasty guidewire to evaluate the effect of percutaneous coronary angioplasty on coronary artery flow velocity.Zhonghua Yi Xue Za Zhi (Taipei). 1995 Mar;55(3):235-41. Zhonghua Yi Xue Za Zhi (Taipei). 1995. PMID: 7780880
-
Practical coronary physiology. Clinical application of the Doppler flow velocity guide wire.Cardiol Clin. 1997 Feb;15(1):77-99. doi: 10.1016/s0733-8651(05)70320-9. Cardiol Clin. 1997. PMID: 9085754 Review.
Cited by
-
Computational fluid dynamics simulations of contrast agent bolus dispersion in a coronary bifurcation: impact on MRI-based quantification of myocardial perfusion.Comput Math Methods Med. 2013;2013:513187. doi: 10.1155/2013/513187. Epub 2013 Feb 28. Comput Math Methods Med. 2013. PMID: 23533541 Free PMC article.
-
Estimation of coronary flow reserve by intracoronary administration of nicorandil: comparison with intracoronary administration of papaverine.Heart Vessels. 1998;13(5):229-36. doi: 10.1007/BF03257245. Heart Vessels. 1998. PMID: 10483772
-
Effect of stenting on coronary flow velocity reserve: comparison of coil and tubular stents.Heart. 1999 Oct;82(4):465-70. doi: 10.1136/hrt.82.4.465. Heart. 1999. PMID: 10490562 Free PMC article.
-
Accurate measurement of pulsatile flow velocity in a small tube phantom: comparison of phase-contrast cine magnetic resonance imaging and intraluminal Doppler guidewire.Jpn J Radiol. 2010 Oct;28(8):571-7. doi: 10.1007/s11604-010-0472-7. Epub 2010 Oct 24. Jpn J Radiol. 2010. PMID: 20972856
-
Hemodynamic diagnostics of epicardial coronary stenoses: in-vitro experimental and computational study.Biomed Eng Online. 2008 Aug 27;7:24. doi: 10.1186/1475-925X-7-24. Biomed Eng Online. 2008. PMID: 18752683 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources