The stentless bioprosthesis: surgical challenges and implications for long-term durability
- PMID: 1389277
The stentless bioprosthesis: surgical challenges and implications for long-term durability
Abstract
An attempt is made to analyse the factors which are expected to influence clinical results following implantation of a stentless porcine bioprosthesis. Long experience with implantation of allograft aortic valves provides a meaningful basis for comparison with a glutaraldehyde-fixed device. Morphological differences between the two valves involve the aortic wall and muscle shelf, and differences in valve preparation include the strength and stiffness of the aorta and the extensibility of the valve leaflets. As a result of these differences, sizing of the porcine valve is expected to be more critical than the allograft valve and the porcine valve is also expected to be more obstructive. Methods for reducing the obstructive element include the use of a composite aortic valve, a porcine pulmonary valve, or a valve in which the aorta is glutaraldehyde-fixed under pressure while the leaflets are pressure-free. The techniques available for implantation, namely freehand insertion, total root replacement and mini-root replacement, are examined.
Similar articles
-
Anatomy of aortic heart valve leaflets: the influence of glutaraldehyde fixation on function.Eur J Cardiothorac Surg. 1992;6 Suppl 1:S25-32; discussion S33. Eur J Cardiothorac Surg. 1992. PMID: 1389275
-
[Hemodynamic performance of newly developed composite stentless porcine aortic valve: in vitro testing and in vivo experiment with sheep].Zhonghua Yi Xue Za Zhi. 2008 Jul 29;88(29):2059-63. Zhonghua Yi Xue Za Zhi. 2008. PMID: 19080436 Chinese.
-
The Toronto root stentless valve in the subcoronary position is hemodynamically superior to the mosaic stented completely supra-annular bioprosthesis.J Heart Valve Dis. 2005 Nov;14(6):814-21; discussion 821. J Heart Valve Dis. 2005. PMID: 16359064
-
Heart valve bioprosthesis durability: a challenge to the new generation of porcine valves.Eur J Cardiothorac Surg. 1992;6 Suppl 1:S82-90. Eur J Cardiothorac Surg. 1992. PMID: 1389286 Review.
-
Porcine aortic valve bioprostheses: morphologic and functional considerations.J Long Term Eff Med Implants. 1992;2(2-3):99-112. J Long Term Eff Med Implants. 1992. PMID: 10148319 Review.
Cited by
-
The use of a stentless porcine bioprosthesis to repair an ascending aortic aneurysm in combination with aortic valve regurgitation.Tex Heart Inst J. 1999;26(3):195-7. Tex Heart Inst J. 1999. PMID: 10524742 Free PMC article.
-
Functional collagen fiber architecture of the pulmonary heart valve cusp.Ann Thorac Surg. 2009 Apr;87(4):1240-9. doi: 10.1016/j.athoracsur.2008.12.049. Ann Thorac Surg. 2009. PMID: 19324159 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Other Literature Sources
Research Materials