Detection of myocardial ischaemia by transthoracic leads in ambulatory electrocardiographic monitoring
- PMID: 1389760
- PMCID: PMC1025072
- DOI: 10.1136/hrt.68.9.286
Detection of myocardial ischaemia by transthoracic leads in ambulatory electrocardiographic monitoring
Abstract
Objective: To determine the best sites for ambulatory monitoring leads to detect myocardial ischaemia.
Patients: 50 consecutive patients recovering from myocardial infarction. Six patients were excluded because of unsatisfactory recordings or baseline electrocardiographic abnormalities that influenced the diagnostic accuracy of ST segment depression. In 38 patients important ST segment changes were seen before the study recordings.
Main outcome measure: Reproducibility of detecting the electrocardiographic ST segment changes with 12 bipolar leads alone or in combination.
Results: The highest reproducibility rate was found in infarcts involving both the anterior and inferior left ventricular walls (80%). The reproducibility decreased as the extent of ventricular wall involvement decreased and was lowest in inferior infarcts (31%) (p < 0.001). For large infarcts the detection rate was almost equal for the 12 study leads, whereas disparity between leads increased as the infarct size decreased. The highest overall reproducibility was found in a transthoracic lead (V2, V9R) (76%). This lead was significantly better (p = 0.03) than lead CM5 (50%). When the transthoracic lead was combined with an inferior lead, the reproducibility increased (82%) and was significantly better than the combination of CM5 and an inferior lead (58%) (p = 0.02).
Conclusions: Extensive ischaemic electrocardiographic changes are better detected than smaller ones and anterior infarcts better than inferior. A transthoracic lead (V2, V9R) was significantly better than CM5 both alone and when CM5 and the transthoracic lead were combined with an inferior lead.
Similar articles
-
Usefulness of a third Holter lead for detection of myocardial ischemia.Am J Cardiol. 1994 Dec 15;74(12):1216-9. doi: 10.1016/0002-9149(94)90551-7. Am J Cardiol. 1994. PMID: 7977093 Clinical Trial.
-
Value of the bipolar lead CM5 in electrocardiography.Br Heart J. 1986 Oct;56(4):372-6. doi: 10.1136/hrt.56.4.372. Br Heart J. 1986. PMID: 3768217 Free PMC article.
-
Value of a bipolar modified inferior lead in detection of inferior myocardial ischaemia.Br Heart J. 1988 Oct;60(4):287-9. doi: 10.1136/hrt.60.4.287. Br Heart J. 1988. PMID: 3190957 Free PMC article.
-
[Isolated ST segment depression from V2 to V4 leads, an early electrocardiographic sign of posterior myocardial infarction].Arch Mal Coeur Vaiss. 1991 Dec;84(12):1815-9. Arch Mal Coeur Vaiss. 1991. PMID: 1793318 Review. French.
-
The role of the initial 12-lead ECG in risk stratification of patients with acute coronary syndrome.Bratisl Lek Listy. 2001;102(9):406-11. Bratisl Lek Listy. 2001. PMID: 11763676 Review.
Cited by
-
Optimum lead positioning for recording bipolar atrial electrocardiograms during sinus rhythm and atrial fibrillation.Clin Cardiol. 1998 Nov;21(11):825-30. doi: 10.1002/clc.4960211108. Clin Cardiol. 1998. PMID: 9825195 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical