. 1992 Sep;65(777):807-13.
doi: 10.1259/0007-1285-65-777-807.

Errors due to non-uniform distribution of fat in dual X-ray absorptiometry of the lumbar spine

Affiliations

Errors due to non-uniform distribution of fat in dual X-ray absorptiometry of the lumbar spine

P Tothill et al. Br J Radiol. 1992 Sep.

Abstract

Errors in spinal dual X-ray absorptiometry (DXA) were studied by analysing X-ray CT scans taken for diagnostic purposes on 20 patients representing a wide range of fat content. The mean difference between the fat thickness over the vertebral bodies and that over a background area in antero-posterior (AP) scanning was 6.7 +/- 8.1 mm for men and 13.4 +/- 4.7 mm for women. For lateral scanning the mean fat thickness difference was only 4 mm for both sexes, but the dispersion was greater, with a standard deviation of 15 mm. To relate these differences to errors in bone mineral, measurements were made of the bone mineral equivalence of fat-equivalent materials on DXA machines from three manufacturers. 10 mm of adipose tissue was equivalent to -0.043 g/cm2 of hydroxyapatite. For AP scanning a non-uniform fat distribution leads to a mean overestimate of 0.029 g/cm2 for men and 0.057 g/cm2 for women. The error exceeded 0.1 g/cm2 in 10% of slices. For lateral scanning the error exceeded 0.1 g/cm2 (about 15% of normal) in a quarter of slices.

PubMed Disclaimer

LinkOut - more resources