Does radial artery pressure accurately reflect aortic pressure?
- PMID: 1395767
- DOI: 10.1378/chest.102.4.1193
Does radial artery pressure accurately reflect aortic pressure?
Abstract
Study objective: Our objective was to determine whether the systolic, diastolic, and mean arterial pressures measured in the radial artery accurately reflect corresponding pressures in the ascending aorta in narcotic-anesthetized patients with known obstructive coronary artery disease, before being subjected to cardiopulmonary bypass (CPB).
Design: This was a prospective study.
Setting: The cardiac operating room of a large, tertiary-care university medical center.
Participants: Fifty-one patients (45 men and six women; age range, 48 to 77 years) with documented atherosclerotic coronary artery disease were studied. All patients underwent elective coronary artery bypass grafting after the study.
Interventions: Patients were premedicated with lorazepam and morphine 60 min before administration of Fentanyl-pancuronium anesthesia. The radial artery was cannulated before induction of anesthesia and the aorta approximately 45 min later. Comparisons of radial and aortic pressures were then performed.
Measurements and results: Radial and aortic pressures were recorded through standard, fluid-filled, high-pressure, 91-cm (36-in) long tubing and disposable transducers, meticulously cleared of air bubbles. Additional measurements included cardiac output, central venous pressure, core temperature, blood gas levels, and hematocrit reading. Radial-aortic pressure differences were as follows: systolic arterial pressure (SAP), 12 +/- 1 mm Hg; mean arterial pressure (MAP), -0.8 +/- 0.3 mm Hg; and diastolic arterial pressure (DAP), -1.0 +/- 0.3 mm Hg. All were significant (p < 0.001), but the SAP difference was more than ten times that of either the MAP or the DAP values. The coefficients of determination (r2) indicated that the radial-aortic dependence was 0.44 for the SAP, 0.90 for the DAP, and 0.98 for the MAP relationship. Plotting the respective differences against the arithmetic mean of simultaneously measured pressures indicated that the radial SAP was 4 to 35 mm Hg higher than the aortic in 42 patients (82 percent) and was 10 to 35 mm Hg higher in 26 patients (51 percent); radial-aortic MAP differences clustered within 3 mm Hg in 47 patients (92 percent); radial DAP was +/- 3 mm Hg different from the aortic in 46 patients (90 percent). The largest MAP difference was -6 mm Hg in one patient. The largest DAP difference was +/- 5 mm Hg in three patients.
Conclusions: In this group of patients, who were studied before undergoing CPB, the radial SAP gave a poor estimate of that present in the ascending aorta, since in more than 50 percent of the cases, the radial SAP was 10 to 35 mm Hg higher than that in the aorta. The radial MAP and DAP are reliable, since in 90 percent and 92 percent of the patients, respectively, the pressure differences were within +/- 3 mm Hg of those in the aorta.
Similar articles
-
Reliability of the radial arterial pressure during anesthesia. Is wrist compression a possible diagnostic test?Chest. 1994 Jan;105(1):69-75. doi: 10.1378/chest.105.1.69. Chest. 1994. PMID: 8275788
-
Prospective evaluation of a method for estimating ascending aortic pressure from the radial artery pressure waveform.Hypertension. 2001 Oct;38(4):932-7. doi: 10.1161/hy1001.096106. Hypertension. 2001. PMID: 11641312
-
Systolic pressure measurement in the ascending aorta: augmentation at the aortic cannula sideport.J Cardiothorac Anesth. 1990 Feb;4(1):25-9. doi: 10.1016/0888-6296(90)90442-i. J Cardiothorac Anesth. 1990. PMID: 2131851
-
A systematic review of invasive, high-fidelity pressure studies documenting the amplification of blood pressure from the aorta to the brachial and radial arteries.J Clin Monit Comput. 2021 Dec;35(6):1245-1252. doi: 10.1007/s10877-020-00599-4. Epub 2020 Oct 9. J Clin Monit Comput. 2021. PMID: 33037525
-
Pulse wave analysis in the assessment of patients with left ventricular assist device.J Heart Lung Transplant. 2001 Jan;20(1):98-102. doi: 10.1016/s1053-2498(00)00177-7. J Heart Lung Transplant. 2001. PMID: 11166618 Review.
Cited by
-
Minimally invasive beat-by-beat monitoring of cardiac power in normal hearts and during acute ventricular dysfunction.Physiol Rep. 2016 Oct;4(19):e12989. doi: 10.14814/phy2.12989. Physiol Rep. 2016. PMID: 27702881 Free PMC article.
-
Carotid characteristics of black South Africans with five-year sustained hypertension.Cardiovasc J Afr. 2016 Jul/Aug;27(4):262-269. doi: 10.5830/CVJA-2016-059. Cardiovasc J Afr. 2016. PMID: 27841914 Free PMC article.
-
Central Hemodynamics for Management of Arteriosclerotic Diseases.J Atheroscler Thromb. 2017 Aug 1;24(8):765-778. doi: 10.5551/jat.40717. Epub 2017 Jun 10. J Atheroscler Thromb. 2017. PMID: 28603219 Free PMC article. Review.
-
Central blood pressure in children and adolescents: non-invasive development and testing of novel transfer functions.J Hum Hypertens. 2017 Dec;31(12):831-837. doi: 10.1038/jhh.2017.59. Epub 2017 Aug 24. J Hum Hypertens. 2017. PMID: 29115295
-
Understanding ventriculo-arterial coupling.Ann Transl Med. 2020 Jun;8(12):795. doi: 10.21037/atm.2020.04.10. Ann Transl Med. 2020. PMID: 32647720 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous