Comparative Study
. 1992 Mar;25(2):67-81.
doi: 10.1111/j.1365-2591.1992.tb00739.x.

Comparison of six files to prepare simulated root canals. 2

Affiliations
Comparative Study

Comparison of six files to prepare simulated root canals. 2

M A al-Omari et al. Int Endod J. 1992 Mar.

Abstract

A total of 300 simulated root canals of various angles and positions of curvature in clear resin blocks were prepared by hand using either K-files, K-Flex files, Flexofiles, Flex-R files, Hedstrom files or Unifiles. Each file type was used to prepare 50 canals employing a linear filing motion and an anticurvature stepback technique. Part 2 of this two-part report describes the efficacy of the files in terms of the prevalence of canal aberrations, the amount and direction of canal transportation and the overall postoperative canal shape. Hourglass shaped canals were found in 4.5% of curved canals, the majority of these being created by K-files and K-Flex files. Perforations were seen in 6.3% of curved canals, the majority being produced by Flex-R files and K-files. Excessive removal of material from along the inner aspect of the canal curve occurred in 10.4% of canals, by far the majority being created by Hedstrom files. Transportation of canals was a consistent finding with all instruments. Two-way analysis of variance revealed how significant differences in the absolute magnitude of transportation occurred between instruments, between canal types and by an interaction between instrument and canal type. In general, transportation was towards the outer aspect of the canal curve at the end-point of preparation and the zip and elbow, but towards the inner aspect at the beginning of the curve. Further towards the orifice, transportation reverted to the outer aspect of the canal curve.(ABSTRACT TRUNCATED AT 250 WORDS)

PubMed Disclaimer

Publication types

MeSH terms

LinkOut - more resources