Percutaneous polymeric stents in porcine coronary arteries. Initial experience with polyethylene terephthalate stents
- PMID: 1423971
- DOI: 10.1161/01.cir.86.5.1596
Percutaneous polymeric stents in porcine coronary arteries. Initial experience with polyethylene terephthalate stents
Abstract
Background: To test the feasibility of percutaneous deployment of intracoronary polymeric stents, a prototype polyethylene terephthalate (PET) stent and a catheter-based delivery system were developed.
Methods and results: Polymeric stents were deployed in the coronary arteries of 11 Yucatan swine: six stents were placed in the left anterior descending coronary artery, four stents were placed in the circumflex artery, and one stent was placed in the right coronary artery. Stent deployment was achieved by withdrawal of an outer delivery sheath, thus allowing the PET stent to self-expand to a preformed configuration. Two animals died during surgery, one during stent placement and the other several hours after implantation due to intracoronary thrombus formation. Two animals were electively sacrificed within 24 hours of stent implant to examine the adequacy of stent deployment within the coronary vessel. The remaining seven animals survived until the termination of the study 4-6 weeks later. Light microscopic examination of the stented vessels showed an extensive neointimal proliferative response with vessel occlusion in all animals who survived initial stent placement. There were two distinct types of histological responses to the PET stent--a chronic foreign body inflammatory response around the stent tines and a neointimal proliferative response in the center of the occluded vessel lumen. The histological response seen in the central area of the vessel was morphologically similar to that seen in patients with restenosis after successful percutaneous transluminal coronary angioplasty, whereas the morphological response seen at the periphery of the stent tine was similar to that exhibited by a chronic foreign body reaction and was not typical of that seen in a restenosis lesion. A ventricular aneurysm also developed in the area of myocardium that was previously supplied by the occluded vessel.
Conclusions: This study demonstrates that percutaneous deployment of polymeric stents in the coronary arteries is technically feasible. The use of PET polymer was associated with an intense proliferative neointimal response that resulted in complete vessel occlusion. Histological examination of the stented segments of the vessel revealed no evidence that dissection of the vessel wall had occurred at the time of initial stent deployment. Although the PET polymer was of similar quality to that used in the manufacture of balloon angioplasty catheters, a toxic chemical or contaminant effect cannot be completely excluded as the stimulus to intimal proliferation. This finding may have relevance to the selection of materials for use as intravascular devices.
Similar articles
-
Coronary stenting with a novel stainless steel balloon-expandable stent: determinants of neointimal formation and changes in arterial geometry after placement in an atherosclerotic model.J Am Coll Cardiol. 1996 Apr;27(5):1270-7. doi: 10.1016/0735-1097(95)00584-6. J Am Coll Cardiol. 1996. PMID: 8609355
-
Restenosis following placement of an intracoronary heparin treated tantulum stent in the hyperlipidemic miniature swine model.J Invasive Cardiol. 1995 Jul-Aug;7(6):173-82. J Invasive Cardiol. 1995. PMID: 10155102
-
Polymeric stenting in the porcine coronary artery model: differential outcome of exogenous fibrin sleeves versus polyurethane-coated stents.J Am Coll Cardiol. 1994 Aug;24(2):525-31. doi: 10.1016/0735-1097(94)90313-1. J Am Coll Cardiol. 1994. PMID: 8034892
-
Comparative pathology: radiation-induced coronary artery disease in man and animals.Semin Interv Cardiol. 1998 Sep-Dec;3(3-4):163-72. Semin Interv Cardiol. 1998. PMID: 10406688 Review.
-
Management of restenosis after coronary intervention.Am Heart J. 1996 Aug;132(2 Pt 1):428-36. doi: 10.1016/s0002-8703(96)90442-1. Am Heart J. 1996. PMID: 8701907 Review.
Cited by
-
Zn-alloy provides a novel platform for mechanically stable bioresorbable vascular stents.PLoS One. 2019 Jan 2;14(1):e0209111. doi: 10.1371/journal.pone.0209111. eCollection 2019. PLoS One. 2019. PMID: 30601854 Free PMC article.
-
Effect of Novel Polymer-Free Nitrogen-Doped Titanium Dioxide Film-Coated Coronary Stent Loaded With Mycophenolic Acid.Front Bioeng Biotechnol. 2021 Oct 29;9:650408. doi: 10.3389/fbioe.2021.650408. eCollection 2021. Front Bioeng Biotechnol. 2021. PMID: 34778222 Free PMC article.
-
Biocompatibility of biodegradable and nonbiodegradable polymer-coated stents implanted in porcine peripheral arteries.Cardiovasc Intervent Radiol. 1995 Jul-Aug;18(4):227-32. doi: 10.1007/BF00239417. Cardiovasc Intervent Radiol. 1995. PMID: 8581902
-
Surface Properties of the Polyethylene Terephthalate (PET) Substrate Modified with the Phospholipid-Polypeptide-Antioxidant Films: Design of Functional Biocoatings.Pharmaceutics. 2022 Dec 15;14(12):2815. doi: 10.3390/pharmaceutics14122815. Pharmaceutics. 2022. PMID: 36559307 Free PMC article.
-
[Local drug delivery and gene therapy].Herz. 1997 Dec;22(6):347-54. doi: 10.1007/BF03044285. Herz. 1997. PMID: 9483440 Review. German.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources