MR-guided focused ultrasound surgery
- PMID: 1430448
- DOI: 10.1097/00004728-199211000-00024
MR-guided focused ultrasound surgery
Abstract
Magnetic resonance guided focused ultrasound surgery provides a minimally invasive controlled method for selectively destroying deep-lying tissue. A thermal analysis of focused ultrasound provides an estimate of the time-dependent temperature distribution and thermal dose required for ultrasound surgery. The temperature distribution is estimated by accumulating heat sources, considering the effects of thermal conductivity, heat content, and perfusion. In this study, both gel phantoms and excised in vitro bovine muscle specimens were imaged in a 1.5 T MR system while heated with a 5 cm diameter, 10 cm focal length, 1.1 MHz transducer. During sonication, the thermal effects were observed with T1-weighted pulse sequences. Below a critical temperature, the heat zone appeared as a dark spot that moved with the focal spot. Above a critical thermal dose, the in vitro tissue was irreversibly altered and the focal lesion was observed on both the MR image and the specimen slice.
Similar articles
-
MR temperature mapping of focused ultrasound surgery.Magn Reson Med. 1994 Jun;31(6):628-36. doi: 10.1002/mrm.1910310608. Magn Reson Med. 1994. PMID: 8057815
-
Magnetic resonance-guided thermal surgery.Magn Reson Med. 1993 Jul;30(1):98-106. doi: 10.1002/mrm.1910300115. Magn Reson Med. 1993. PMID: 8371680
-
Focused US system for MR imaging-guided tumor ablation.Radiology. 1995 Mar;194(3):731-7. doi: 10.1148/radiology.194.3.7862971. Radiology. 1995. PMID: 7862971
-
MRI-controlled ultrasound thermal therapy.IEEE Pulse. 2011 Sep-Oct;2(5):39-47. doi: 10.1109/MPUL.2011.942604. IEEE Pulse. 2011. PMID: 25372968 Review.
-
Magnetic resonance temperature imaging for focused ultrasound surgery: a review.Top Magn Reson Imaging. 2006 Jun;17(3):153-63. doi: 10.1097/RMR.0b013e3180377bc3. Top Magn Reson Imaging. 2006. PMID: 17414072 Review.
Cited by
-
MR thermometry for monitoring tumor ablation.Eur Radiol. 2007 Sep;17(9):2401-10. doi: 10.1007/s00330-007-0646-6. Epub 2007 May 22. Eur Radiol. 2007. PMID: 17701184 Review.
-
Current State of Potential Mechanisms Supporting Low Intensity Focused Ultrasound for Neuromodulation.Front Hum Neurosci. 2022 Apr 25;16:872639. doi: 10.3389/fnhum.2022.872639. eCollection 2022. Front Hum Neurosci. 2022. PMID: 35547195 Free PMC article. Review.
-
SPatiotemporal-ENcoded acoustic radiation force imaging of focused ultrasound.Front Hum Neurosci. 2023 Apr 27;17:1184629. doi: 10.3389/fnhum.2023.1184629. eCollection 2023. Front Hum Neurosci. 2023. PMID: 37180550 Free PMC article.
-
Ex Vivo HIFU Experiments Using a $32 \times 32$ -Element CMUT Array.IEEE Trans Ultrason Ferroelectr Freq Control. 2016 Dec;63(12):2150-2158. doi: 10.1109/TUFFC.2016.2606126. IEEE Trans Ultrason Ferroelectr Freq Control. 2016. PMID: 27913330 Free PMC article.
-
Focused Ultrasound: Noninvasive Image-Guided Therapy.Invest Radiol. 2025 Mar 1;60(3):205-219. doi: 10.1097/RLI.0000000000001116. Epub 2024 Aug 21. Invest Radiol. 2025. PMID: 39163359 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical