Validation of a portable indirect calorimetry system for measurement of energy expenditure in sick preterm infants
- PMID: 1444562
- PMCID: PMC1590467
- DOI: 10.1136/adc.67.10_spec_no.1207
Validation of a portable indirect calorimetry system for measurement of energy expenditure in sick preterm infants
Abstract
A portable indirect calorimeter adapted from adult use was validated for use in preterm infants. Oxygen consumption (VO2) and carbon dioxide production (VCO2) were subsequently measured in 16 preterm infants breathing spontaneously in room air (canopy mode) and in nine preterm infants receiving intermittent positive pressure ventilation (ventilator mode). Validation of the system was performed using a gas injection technique with nitrogen to simulate VO2 and carbon dioxide for VCO2. Mean errors in validation of the canopy mode were 1.4% and 0.2% for VO2 and VCO2 with limits of agreement of 0.6 (+2SD) ml/min and -1.3 (-2SD) ml/min, and 0.9 (+2SD) ml/min and -2.3 (-2SD) ml/min respectively. In validation of the ventilator mode mean errors were -1.8% and -5.05% for VO2 and VCO2 with limits of agreement of 1.02 (+2SD) ml/min and -0.74 (-2SD) ml/min, and 0.93 (+2SD) ml/min and -1.45 (-2SD) ml/min respectively. Values of VO2 and VCO2 in 16 preterm infants in the canopy mode were 6.2 ml/kg/min (0.5 1SD) and 6.7 ml/kg/min (0.6 1SD) and in nine preterm infants in the ventilator mode 4.98 ml/kg/min (1.09 1SD) and 4.74 ml/min/kg (1.08 1SD) respectively. Mean energy expenditure was 45.5 kcal (191 kJ)kg/day for infants measured in the canopy mode and 35.5 kcal (149 kJ)/kg/day for ventilated infants. This metabolic system can be adapted for use in the newborn but accuracy is reduced when it is used in those weighing less than 1000 g.
Similar articles
-
Comparison of face mask, head hood, and canopy for breath sampling in flow-through indirect calorimetry to measure oxygen consumption and carbon dioxide production of preterm infants < 1500 grams.Pediatr Res. 1997 Jan;41(1):139-44. doi: 10.1203/00006450-199701000-00022. Pediatr Res. 1997. PMID: 8979303
-
A water-sealed indirect calorimeter for measurement of oxygen consumption (VO2), carbon dioxide production (VCO2), and energy expenditure in infants.JPEN J Parenter Enteral Nutr. 1988 May-Jun;12(3):256-9. doi: 10.1177/0148607188012003256. JPEN J Parenter Enteral Nutr. 1988. PMID: 3134559
-
Evaluation of the accuracy and precision of a new generation indirect calorimeter in canopy dilution mode.Clin Nutr. 2020 Jun;39(6):1927-1934. doi: 10.1016/j.clnu.2019.08.017. Epub 2019 Sep 10. Clin Nutr. 2020. PMID: 31543335
-
Indirect calorimetry: technical aspects.J Am Diet Assoc. 1997 Oct;97(10 Suppl 2):S154-60. doi: 10.1016/s0002-8223(97)00754-2. J Am Diet Assoc. 1997. PMID: 9336580 Review.
-
Indirect calorimetry: a trend toward continuous metabolic assessment.AACN Clin Issues. 2003 May;14(2):155-67; quiz 266. doi: 10.1097/00044067-200305000-00005. AACN Clin Issues. 2003. PMID: 12819453 Review.
Cited by
-
Low birthweight infants and total parenteral nutrition immediately after birth. III. Randomised study of energy substrate utilisation, nitrogen balance, and carbon dioxide production.Arch Dis Child Fetal Neonatal Ed. 1995 Jul;73(1):F13-6. doi: 10.1136/fn.73.1.f13. Arch Dis Child Fetal Neonatal Ed. 1995. PMID: 7552589 Free PMC article. Clinical Trial.
-
Measurement of carbon dioxide production in very low birth weight babies.Arch Dis Child Fetal Neonatal Ed. 2000 Jul;83(1):F50-5. doi: 10.1136/fn.83.1.f50. Arch Dis Child Fetal Neonatal Ed. 2000. PMID: 10873173 Free PMC article.
-
Validation of indirect calorimetry for measurement of energy expenditure in healthy volunteers undergoing pressure controlled non-invasive ventilation support.J Clin Monit Comput. 2012 Feb;26(1):37-43. doi: 10.1007/s10877-011-9331-z. Epub 2011 Dec 30. J Clin Monit Comput. 2012. PMID: 22207315
-
Evolution of Resting Energy Expenditure, Respiratory Quotient, and Adiposity in Infants Recovering from Corrective Surgery of Major Congenital Gastrointestinal Tract Anomalies: A Cohort Study.Nutrients. 2020 Oct 11;12(10):3093. doi: 10.3390/nu12103093. Nutrients. 2020. PMID: 33050623 Free PMC article.
-
Predictive equations over-estimate the resting energy expenditure in amyotrophic lateral sclerosis patients who are dependent on invasive ventilation support.Nutr Metab (Lond). 2010 Aug 26;7:70. doi: 10.1186/1743-7075-7-70. Nutr Metab (Lond). 2010. PMID: 20796286 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources