Comparison of unipolar and bipolar ventricular paced evoked responses
- PMID: 1449924
- PMCID: PMC1025140
- DOI: 10.1136/hrt.68.10.398
Comparison of unipolar and bipolar ventricular paced evoked responses
Abstract
Objectives: To study the differences between endocardial bipolar and unipolar ventricular paced evoked responses and surface electrocardiograms.
Patients: 10 patients with conduction system disease awaiting insertion of a permanent pacemaker were studied with temporary ventricular pacing from the right ventricular apex.
Main outcome measure: Comparison of the durations of the QRS complexes and QTa and QTe intervals of the endocardial bipolar paced evoked response and the surface electrocardiogram with those of the reference unipolar paced evoked response.
Results: By comparison with the unipolar reference, the mean durations of the QRS complexes of the bipolar signal and the surface electrocardiogram were 41.8% and 132.1% respectively. The mean QTa interval was 85.9% and 112.2% respectively and the mean QTe interval was 86.9% and 109.5% respectively. All these differences were significant. The amplitudes of the unipolar QRS complexes and T waves were significantly larger than those recorded in the bipolar configuration.
Conclusions: Differences between the unipolar and bipolar ventricular paced evoked responses are significant. The time course of the unipolar signal is closer to that of the surface electrocardiogram. This indicates that the unipolar paced evoked response does not reflect local electrophysiological events, as has been suggested previously.
Similar articles
-
A comparison of QRS complexes resulting from unipolar and bipolar pacing: implications for pace-mapping.Pacing Clin Electrophysiol. 1991 May;14(5 Pt 1):823-32. doi: 10.1111/j.1540-8159.1991.tb04113.x. Pacing Clin Electrophysiol. 1991. PMID: 1712960
-
Electrocardiographic features of atrial depolarization during pacing of right atrial appendage.Indian Heart J. 1991 Mar-Apr;43(2):79-82. Indian Heart J. 1991. PMID: 1752619
-
Uni- and bipolar pacing threshold measurements with capturecontrol, a new automatic pacemaker function for capture verification. Comparison of the automatic and the manual pacing threshold measurement.Cardiology. 1999;92(3):210-3. doi: 10.1159/000006973. Cardiology. 1999. PMID: 10754353 Clinical Trial.
-
Analysis of the morphology of the unipolar endocardial paced evoked response.Pacing Clin Electrophysiol. 1990 Mar;13(3):302-13. doi: 10.1111/j.1540-8159.1990.tb02044.x. Pacing Clin Electrophysiol. 1990. PMID: 1690403 Review.
-
[Indications spectrum for temporary and permanent cardiac pacemaker therapy].Z Gerontol. 1994 Sep-Oct;27(5):337-40. Z Gerontol. 1994. PMID: 7810202 Review. German.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous