Case Reports
. 1992 Nov;39(9):997-9.
doi: 10.1007/BF03008353.

Inspiratory valve malfunction in a circle system: pitfalls in capnography

Affiliations
Case Reports

Inspiratory valve malfunction in a circle system: pitfalls in capnography

A Y Kumar et al. Can J Anaesth. 1992 Nov.

Abstract

Capnography is a useful technique in monitoring the integrity of anaesthetic equipment such as the malfunctioning of unidirectional valves in circle system. However, the lack of a precise mechanism in existing capnographs to identify the start of inspiration and the beginning of expiration in the capnograms, makes the analysis of the carbon dioxide waveforms during inspiration difficult and thus results in inaccurate assessment of rebreathing. We report a case where, during the malfunction of the inspiratory unidirectional valve in the circle system, the capnograph failed to detect the presence of substantial rebreathing. Critical analysis of the capnogram recorded during the malfunction revealed that there was substantial rebreathing which was underestimated by the capnograph as it reports only the lowest CO2 concentration rebreathed during inspiration in such abnormal situations.

PubMed Disclaimer

References

    1. J Clin Monit. 1990 Jul;6(3):217-25 - PubMed
    1. J Clin Monit. 1987 Jul;3(3):194-200 - PubMed
    1. Can J Anaesth. 1988 Sep;35(5):473-5 - PubMed
    1. Anesth Analg. 1984 May;63(5):536-7 - PubMed

Publication types