Measurement of diaphragm loading during pressure support ventilation

Affiliations

Measurement of diaphragm loading during pressure support ventilation

Brigitte Fauroux et al. Intensive Care Med. 2003 Nov.

Abstract

Objective: The diaphragmatic pressure-time product (PTPdi) has been used to quantify the loading and unloading of the diaphragm. The validity of the relationship between PTPdi and diaphragm electrical activity (EMGdi) during pressure-support ventilation (PSV) is unclear. We examined this relationship.

Design and setting: Physiological study in a physiology laboratory.

Subjects: Six healthy adults.

Interventions: Spontaneous breathing (SB) and two levels of PSV (6 and 12 cmH(2)O), breathing room air and incremental concentrations of carbon dioxide, sufficient to achieve an EMGdi signal of approximately 200% of baseline value.

Measurements and results: We measured the electrical (EMGdi) and mechanical (PTPdi) activity of the diaphragm using oesophageal electrode and oesophageal and gastric balloon catheters. The relationship between EMGdi and PTPdi during SB was linear in five subjects and curvilinear in one. However, with PSV 12 cmH(2)O we observed that the relationship between EMGdi and PTPdi was 'left shifted'; specifically, for any given level of EMGdi the PTPdi was smaller with PSV 12 cmH(2)O than during SB. However, when PTPdi was converted to power (the product of pressure and flow) the tendency to left shift was largely reversed.

Conclusions: We conclude that when assessing of diaphragm unloading during PSV flow measurements are required. Where flow is constant, PTPdi is a valid measure of diaphragm unloading, but if not these data may be used to make an appropriate correction.

PubMed Disclaimer

References

    1. Am Rev Respir Dis. 1991 Mar;143(3):469-75 - PubMed
    1. Am J Respir Crit Care Med. 2001 Aug 1;164(3):419-24 - PubMed
    1. Nat Med. 1999 Dec;5(12):1433-6 - PubMed
    1. Intensive Care Med. 1996 Nov;22(11):1147-54 - PubMed
    1. Intensive Care Med. 1995 Nov;21(11):871-9 - PubMed

Publication types

MeSH terms

LinkOut - more resources