Angular dependence of the luminance and contrast in medical monochrome liquid crystal displays
- PMID: 14596296
- DOI: 10.1118/1.1606449
Angular dependence of the luminance and contrast in medical monochrome liquid crystal displays
Abstract
Active-matrix liquid crystal displays (AMLCDs) are light-modulating devices that generate images by differentially transmitting a nearly uniform luminous field provided by a backlight. While emissive displays exhibit a quasi-Lambertian emission with almost constant contrast at off-normal viewing, the anisotropy of the electro-optic effect that controls light transmission in AMLCDs causes a pixel luminance that varies, sometimes strongly, with viewing angle. These variations are not identical for all gray levels and can eventually cause grayscale inversions. In this paper, we measured the luminance emission of a monochrome medical AMLCD, a medical cathode-ray tube monitor, and a color desktop AMLCD, using a collimated photopic probe positioned on a manual rotation arm, and a research radiometer with automatic readout. The probe measures luminance with a small acceptance angle and provides optical shielding from emissions at other viewing directions that contaminate the readings. We obtained luminance response curves versus angle in the vertical, horizontal and at 45 degrees diagonal directions. The display systems were calibrated to reflect the DICOM Part 3.14 standard grayscale display function (GDF) when measured using the manufacturer's probe and software tools. We analyzed the measurements at different viewing directions with respect to their departure from the GDF by computing the normalized contrast (deltaL/L) as a function of the DICOM just-noticeable difference index. Although cathode-ray tubes are known to be quasi-Lambertian emitters, the luminance at normal viewing is higher than the luminance observed at large angles. This decrease in luminance is however proportionally similar for all gray levels, resulting in a relatively flat contrast response for all angles. In addition to being more pronounced, the angular variation in AMLCDs does not follow the same profile at different intensities with the subsequent variation in the achieved display contrast. The changes due to off-normal viewing are substantial at large angles in the horizontal and vertical directions, and much worse in the diagonal viewing directions.
Similar articles
-
Effect of viewing angle on luminance and contrast for a five-million-pixel monochrome display and a nine-million-pixel color liquid crystal display.J Digit Imaging. 2004 Dec;17(4):264-70. doi: 10.1007/s10278-004-1021-7. J Digit Imaging. 2004. PMID: 15692870 Free PMC article.
-
Viewing angle performance of medical liquid crystal displays.Med Phys. 2006 Mar;33(3):645-54. doi: 10.1118/1.2168430. Med Phys. 2006. PMID: 16878568
-
Goniometric and conoscopic measurements of angular display contrast for one-, three-, five-, and nine-million-pixel medical liquid crystal displays.Med Phys. 2004 Dec;31(12):3452-60. doi: 10.1118/1.1824198. Med Phys. 2004. PMID: 15651628
-
High-fidelity electronic display of digital radiographs.Radiographics. 1999 Nov-Dec;19(6):1653-69. doi: 10.1148/radiographics.19.6.g99no081653. Radiographics. 1999. PMID: 10555680 Review.
-
AAPM/RSNA tutorial on equipment selection: PACS equipment overview: display systems.Radiographics. 2004 May-Jun;24(3):879-89. doi: 10.1148/rg.243035133. Radiographics. 2004. PMID: 15143237 Review.
Cited by
-
ACR-AAPM-SIIM practice guideline for determinants of image quality in digital mammography.J Digit Imaging. 2013 Feb;26(1):10-25. doi: 10.1007/s10278-012-9521-3. J Digit Imaging. 2013. PMID: 22992865 Free PMC article. No abstract available.
-
Visual assessment of angular response in medical liquid crystal displays.J Digit Imaging. 2006 Sep;19(3):240-8. doi: 10.1007/s10278-006-0633-5. J Digit Imaging. 2006. PMID: 16741662 Free PMC article.
-
Increasing the number of gray shades in medical display systems--how much is enough?J Digit Imaging. 2007 Dec;20(4):422-32. doi: 10.1007/s10278-006-1052-3. J Digit Imaging. 2007. PMID: 17195900 Free PMC article.
-
Differential use of image enhancement techniques by experienced and inexperienced observers.J Digit Imaging. 2005 Dec;18(4):311-5. doi: 10.1007/s10278-005-7666-z. J Digit Imaging. 2005. PMID: 16142436 Free PMC article.
-
Assessment of PACS display systems.J Digit Imaging. 2005 Dec;18(4):287-95. doi: 10.1007/s10278-005-6974-7. J Digit Imaging. 2005. PMID: 16094505 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources